HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2017-00120
- Event Type
- Malfunction
- Date Received
- March 2, 2017
- Date of Event
- January 31, 2017
- Report Date
- March 2, 2017
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THERE IS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER DEVICE 16-02-80 IS NOT DISTRIBUTED IN THE USA. HOWEVER, IT IS SIMILAR TO THE HEATER-COOLER DEVICE 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). LIVANOVA IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF LIVANOVA HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). THROUGH FOLLOW-UP COMMUNICATION WITH THE CUSTOMER, LIVANOVA (B)(4) CONFIRMED THAT THERE ARE NO KNOWN PATIENT INFECTIONS RELATED TO THE REPORTED CONTAMINATION. THE FACILITY HAS STOPPED THE USE OF THE INVOLVED DEVICE AND HAS REQUESTED RETURN TO LIVANOVA DEUTSCHLAND FOR DEEP DISINFECTION. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.
LIVANOVA (B)(4) RECEIVED A REPORT THAT MYCOBACTERIUM CHIMAERA WAS CULTURED FROM WATER SAMPLES TAKEN FROM THE HEATER-COOLER SYSTEM 3T DURING MAINTENANCE. THERE IS NO KNOWN PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155266 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |