FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 6372454 · Received March 2, 2017

Report

Report Number
9611109-2017-00121
Event Type
Malfunction
Date Received
March 2, 2017
Date of Event
January 31, 2017
Report Date
March 2, 2017
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Removal / Correction Number
Z-2076/2081-2015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO KNOWN PATIENT INVOLVEMENT. PMA 510(K): THE HEATER-COOLER DEVICE 16-02-80 IS NOT DISTRIBUTED IN THE USA. HOWEVER, IT IS SIMILAR TO THE HEATER-COOLER DEVICE 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). RECALL NUMBER: LIVANOVA IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF LIVANOVA HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). THROUGH FOLLOW-UP COMMUNICATION WITH THE CUSTOMER, LIVANOVA (B)(4) CONFIRMED THAT THERE ARE NO KNOWN PATIENT INFECTIONS RELATED TO THE REPORTED CONTAMINATION. THE FACILITY HAS STOPPED THE USE OF THE INVOLVED DEVICE AND HAS REQUESTED RETURN TO LIVANOVA DEUTSCHLAND FOR DEEP DISINFECTION. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT MYCOBACTERIUM CHIMAERA WAS CULTURED FROM WATER SAMPLES TAKEN FROM THE HEATER-COOLER SYSTEM 3T DURING MAINTENANCE. THERE IS NO KNOWN PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155265 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1