FDA Adverse Event Injury Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 6368678 · Received March 1, 2017

Report

Report Number
9611109-2017-00112
Event Type
Injury
Date Received
March 1, 2017
Date of Event
November 20, 2015
Report Date
March 1, 2017
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Removal / Correction Number
Z-2076/2081-2015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SERIAL NUMBER IS UNKNOWN, AS THE FACILITY DOES NOT TRACK WHICH UNIT IS USED FOR A GIVEN PROCEDURE. THE HEATER-COOLER DEVICE 16-02-80 IS NOT DISTRIBUTED IN THE USA. HOWEVER, IT IS SIMILAR TO THE HEATER-COOLER DEVICE 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K052601). AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE CANNOT BE DETERMINED. LIVANOVA IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF LIVANOVA HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). THE REPORTING FACILITY (PRINCE OF WALES PRIVATE HOSPITAL) HAS REPORTED THAT MULTIPLE HEATER-COOLER DEVICES HAVE BEEN FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHIMAERA (SEE MEDWATCH REPORTS 9611109-2017-00114, 9611109-2017-00115 AND 9611109-2017-00116 FOR DETAILS ON THE UNITS). THE FACILITY DOES NOT TRACK WHICH UNIT IS USED FOR A GIVEN PROCEDURE, SO IT IS UNKNOWN WHICH OF THESE DEVICES WAS USED DURING THE PATIENT'S PROCEDURE. THE INVESTIGATION IS ON-GOING. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT THAT A PATIENT TESTED POSITIVE FOR MYCOBACTERIUM CHIMAERA INFECTION AFTER UNDERGOING AN AORTIC VALVE REPLACEMENT PROCEDURE THAT UTILIZED A HEATER-COOLER SYSTEM 3T.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153252 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1 80 YR Other