ALARIS® PUMP MODULE ADMINISTRATION SET
Report
- Report Number
- 9616066-2017-00353
- Event Type
- Malfunction
- Date Received
- February 28, 2017
- Report Date
- February 10, 2017
- Manufacturer
- CAREFUSION
- Product Code
- FPA
- PMA / PMN Number
- K944320
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER¿S REPORT OF A HOLE IN THE TUBING WAS CONFIRMED. VISUAL INSPECTION OBSERVED THAT THE SILICONE SEGMENT HAD A TEAR NEAR THE UPPER FITMENT. THE TEAR MEASURED 0.058 INCHES LONG. VISUAL EXAMINATION UNDER MAGNIFICATION NOTED NO CRUSH MARK TO THE UPPER BLUE FITMENT. FUNCTIONAL AND PRESSURE TESTING WAS PERFORMED; A LEAK WAS OBSERVED COMING OUT FROM THE SILICONE TUBING NEAR THE UPPER FITMENT. THE CAUSE OF THE LEAK WAS A TEAR IN THE SILICONE SEGMENT. THE ROOT CAUSE OF THE TEAR WAS NOT IDENTIFIED.
(B)(4), LOT: 16125108 IS 07613203021012. CONCOMITANT MEDICAL PRODUCTS: 250 ML BAXTER BAG NDC 0338-0049-02, LOT Y221309, EXP JUL 18, 09.% SODIUM CHLORIDE INJECTION; THERAPY DATE UNKNOWN THE AFFECTED PRODUCT HAS BEEN RECEIVED AND THE EVALUATION IS PENDING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETED.
THE CUSTOMER REPORTED THAT WHILE PRIMING LEVOPHED THAT WAS CONNECTED TO THE PATIENT, THEY NOTICED FLUID DRIPPING ON THE FLOOR. AFTER LOOKING AT THE TUBING THEY FOUND A HOLE IN THE TUBING AND THE MEDICATION WAS LEAKING FROM IT. THERE IS NO REPORT OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149082 | ALARIS® PUMP MODULE ADMINISTRATION SET | SET, ADMINISTRATION, INTRAVASCULAR | FPA | CAREFUSION | 2420-0007 | 16125108 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |