PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2017-07904
- Event Type
- Malfunction
- Date Received
- February 28, 2017
- Date of Event
- July 13, 2013
- Report Date
- July 13, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD
- Reporter Occupation
- PATIENT
Narratives
UNIT RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO TRACES OF MOISTURE WERE NOTED AT ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. UNIT PASSED THE SELF TEST, A21 ERROR TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. NO UNEXPECTED A21 ERROR ALARMS WERE NOTED. UNIT RECEIVED WITH CRACKED CASE AT KCD WINDOW CORNERS, RESERVOIR TUBE AND BATTERY TUBE THREADS. UNIT RECEIVED WITH SCRATCHED LCD WINDOW. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD UNEXPECTED RESTART ALARM AND BUTTON ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 55 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 146352 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |