FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 6365467 · Received February 28, 2017

Report

Report Number
2032227-2017-07904
Event Type
Malfunction
Date Received
February 28, 2017
Date of Event
July 13, 2013
Report Date
July 13, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

UNIT RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO TRACES OF MOISTURE WERE NOTED AT ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. UNIT PASSED THE SELF TEST, A21 ERROR TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. NO UNEXPECTED A21 ERROR ALARMS WERE NOTED. UNIT RECEIVED WITH CRACKED CASE AT KCD WINDOW CORNERS, RESERVOIR TUBE AND BATTERY TUBE THREADS. UNIT RECEIVED WITH SCRATCHED LCD WINDOW. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON JULY 1, 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD UNEXPECTED RESTART ALARM AND BUTTON ERROR ALARM. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 55 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146352 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAL

Patients

Seq Age Sex Outcome Treatment
1 43 YR