FDA Adverse Event Malfunction Summary report: N

HF-CABLE, MONOPOLAR, 4 M, UES-30/-40,

MDR report key: 6364691 · Received February 28, 2017

Report

Report Number
2951238-2017-00132
Event Type
Malfunction
Date Received
February 28, 2017
Date of Event
January 31, 2017
Report Date
February 28, 2017
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCJ
PMA / PMN Number
PK944201
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR EVALUATION. THE MOST LIKELY CAUSE FOR THE REPORTED EVENT IS DUE TO USER HANDLING. THE INSTRUCTION MANUAL CONTAINS SEVERAL WARNING STATEMENTS IN AN EFFORT TO PREVENT DAMAGE TO THE CABLE. ¿VISUALLY INSPECT THE CABLE AND THE PLUGS FOR IRREGULARITIES ON THE SURFACE. DO NOT USE A CABLE WITH BRITTLE OR DEFECTIVE INSULATION. REPLACE THE CABLE. IN ORDER TO PLUG OR UNPLUG THE CABLE, ALWAYS PULL AT THE PLUG. NEVER PULL AT THE CABLE OR RISK OF DAMAGING THE CABLE MAY OCCUR.¿ ALSO, ¿WHEN USED AS INTENDED THIS PRODUCT IS MORE OR LESS SUBJECT TO WEAR DEPENDING ON THE INTENSITY OF USE. DO NOT USE THE HF CABLE AFTER ONE YEAR OF USE.¿ IF ADDITIONAL INFORMATION BECOMES AVAILABLE OR IF THE DEVICE IS RETURNED FOR EVALUATION, THIS REPORT WILL BE UPDATED ACCORDINGLY.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT THE HF-CABLE SPARKED AND FAILED DURING AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR HF CABLE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148708 HF-CABLE, MONOPOLAR, 4 M, UES-30/-40, HF CABLE GCJ OLYMPUS WINTER & IBE GMBH A0393 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1