22 G X 1 1/2 IN. BD ECLIPSE¿ HYPODERMIC NEEDLE
Report
- Report Number
- 8041187-2017-00024
- Event Type
- Malfunction
- Date Received
- February 27, 2017
- Date of Event
- February 8, 2017
- Report Date
- February 23, 2017
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FMI
- PMA / PMN Number
- K161170
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED. RESULTS: A SAMPLE IS NOT AVAILABLE FOR EVALUATION. HOWEVER, THE CUSTOMER PROVIDED A PHOTO OF THE AFFECTED DEVICE FOR ANALYSIS. A PHOTO INSPECTION SHOWED THE SAFETY SHIELD DISENGAGED FROM THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 5323003. A MANUFACTURING REVIEW REVEALED NO ABNORMALITIES WITH PREVENTATIVE MAINTENANCE, CALIBRATION, OR EQUIPMENT THAT COULD HAVE INFLUENCED THE CUSTOMER'S REPORTED ISSUE. CONCLUSION: ALTHOUGH THE CUSTOMER'S PHOTO SHOWED THE INDICATED DEFECT, WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE. REMEDIAL ACTION REQUIRED: CAPAS (B)(4) HAVE BEEN OPENED TO IDENTIFY AND ADDRESS THE POTENTIAL CAUSES OF SAFETY SHIELD DISENGAGEMENT. ADDITIONALLY, FIELD ACTION NOTIFICATION MSS-16-837-FA WAS INITIATED AND A PRODUCT ADVISORY LETTER WAS SENT ON 12/29/2016. (B)(4).
IT WAS REPORTED THAT AFTER A NURSE HAD GIVEN A VITAMIN B12 INJECTION WITH A 22 G X 1 1/2 IN. BD ECLIPSE¿ HYPODERMIC NEEDLE, HE/SHE ACTIVATED THE SAFETY MECHANISM ON A COUNTER TOP AND THE SAFETY SHIELD BROKE OFF. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143927 | 22 G X 1 1/2 IN. BD ECLIPSE¿ HYPODERMIC NEEDLE | SAFETY ENGINEERED HYPODERMIC NEEDLE | FMI | BECTON DICKINSON MEDICAL (SINGAPORE) | 5323003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |