Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 182, 181 AND 178 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 95 - 115 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN CABINET. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 01/31/2018 AND OPEN VIAL DATE AT TIME OF CALL IS A WEEK AGO. CUSTOMER STATED HE HAS NOT HAD ANY RECENT CHANGES IN HIS DIET, EXERCISE, OR MEDICATIONS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143999 METRIX AIR BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. METRIX AIR MT1914 00021292007447

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY