MREYE FLIPPER PDA CLOSURE DETACHABLE EMBOLIZATION COIL
Report
- Report Number
- 3002808486-2017-00543
- Event Type
- Injury
- Date Received
- February 27, 2017
- Date of Event
- February 7, 2013
- Report Date
- February 15, 2013
- Manufacturer
- WILLIAM COOK EUROPE
- Product Code
- KRD
- UDI-DI
- 00827002202368
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). SIMILAR TO DEVICE UNDER 510(K) K063619 (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: THE USED IMWCE-PDA COIL AND DELIVERY WIRE WERE RETURNED TO AID THE INVESTIGATION. VISUAL INSPECTION SHOWED THE PROXIMAL PART OF THE COIL WAS UNRAVELED. HOWEVER, NO UNRAVELING OF THE THREAD OF THE COIL WAS OBSERVED. INSPECTION OF DELIVERY WIRE FOUND THREAD COMPLETELY UNRAVELED. IT IS ASSUMED THAT COIL DETACHED UNINTENTIONALLY AND UNRAVELING OF THE DELIVERY WIRE OCCURRED AFTER DETACHMENT. THE FINDINGS INDICATE THAT EXCESSIVE FORCE WAS APPLIED TO THE DELIVERY WIRE LEADING IT TO UNRAVEL. IT IS ASSUMED THAT THE COIL UNRAVELED DURING REMOVAL FROM THE ARTERY SINCE NO UNRAVELING OF THE THREAD IS OBSERVED. UNINTENDED DETACHMENT OF THE COIL CAN OCCUR IF THE DELIVERY WIRE ACCIDENTALLY IS ROTATED COUNTER CLOCKWISE DURING INSERTION NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE SURGEON LOST THE COIL IN THE PULMONARY ARTERY: WHEN THE COIL BROKE OFF FROM THE SHEATH AND THE GUIDE. THE SPIRES WERE ABNORMALLY LENGTHENED AND DID NOT RESIST THE TRACTION. THE SURGEON RECOVERED THE COIL BY CATHETERISM WITH A LASSO. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143753 | MREYE FLIPPER PDA CLOSURE DETACHABLE EMBOLIZATION COIL | KRD DEVICE, EMBOLIZATION, VASCULAR | KRD | WILLIAM COOK EUROPE | 00827002202368 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |