FDA Adverse Event Injury Summary report: N

MREYE FLIPPER PDA CLOSURE DETACHABLE EMBOLIZATION COIL

MDR report key: 6359912 · Received February 27, 2017

Report

Report Number
3002808486-2017-00543
Event Type
Injury
Date Received
February 27, 2017
Date of Event
February 7, 2013
Report Date
February 15, 2013
Manufacturer
WILLIAM COOK EUROPE
Product Code
KRD
UDI-DI
00827002202368
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SIMILAR TO DEVICE UNDER 510(K) K063619 (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: THE USED IMWCE-PDA COIL AND DELIVERY WIRE WERE RETURNED TO AID THE INVESTIGATION. VISUAL INSPECTION SHOWED THE PROXIMAL PART OF THE COIL WAS UNRAVELED. HOWEVER, NO UNRAVELING OF THE THREAD OF THE COIL WAS OBSERVED. INSPECTION OF DELIVERY WIRE FOUND THREAD COMPLETELY UNRAVELED. IT IS ASSUMED THAT COIL DETACHED UNINTENTIONALLY AND UNRAVELING OF THE DELIVERY WIRE OCCURRED AFTER DETACHMENT. THE FINDINGS INDICATE THAT EXCESSIVE FORCE WAS APPLIED TO THE DELIVERY WIRE LEADING IT TO UNRAVEL. IT IS ASSUMED THAT THE COIL UNRAVELED DURING REMOVAL FROM THE ARTERY SINCE NO UNRAVELING OF THE THREAD IS OBSERVED. UNINTENDED DETACHMENT OF THE COIL CAN OCCUR IF THE DELIVERY WIRE ACCIDENTALLY IS ROTATED COUNTER CLOCKWISE DURING INSERTION NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.

Description of Event or Problem · 1

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE SURGEON LOST THE COIL IN THE PULMONARY ARTERY: WHEN THE COIL BROKE OFF FROM THE SHEATH AND THE GUIDE. THE SPIRES WERE ABNORMALLY LENGTHENED AND DID NOT RESIST THE TRACTION. THE SURGEON RECOVERED THE COIL BY CATHETERISM WITH A LASSO. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143753 MREYE FLIPPER PDA CLOSURE DETACHABLE EMBOLIZATION COIL KRD DEVICE, EMBOLIZATION, VASCULAR KRD WILLIAM COOK EUROPE 00827002202368

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention