RETROGRADE DRL 10MM

MDR report key: 6359497 · Received February 25, 2017

Report

Report Number
1643264-2017-00114
Event Type
Malfunction
Date Received
February 25, 2017
Date of Event
February 15, 2017
Report Date
April 5, 2017
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GFG
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

VISUAL ASSESSMENT OF THE DEVICE CONFIRMED THE REPORTED COMPLAINT OF BREAKAGE. THE DRILL HEAD HAS SPLIT IN TWO PIECES AND HAS COME FREE FROM THE DRILL SHAFT. THE BREAK IS A CLEAN SHEAR WITH A SLIGHT TWIST. THE ACTUATOR WIRE IS SPIRALED IN THE FORWARD DRILLING POSITION.

Description of Event or Problem · 1

IT WAS REPORTED THE RETROGRADE DRILL (10MM) BROKE DURING A PROCEDURE. ALL PIECES REMOVED FROM PATIENT. A BACK UP DEVICE WAS UTILIZED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.