FDA Adverse Event Malfunction Summary report: N

E SERIES DEFIBRILLATOR

MDR report key: 6359458 · Received February 25, 2017

Report

Report Number
1220908-2017-00363
Event Type
Malfunction
Date Received
February 25, 2017
Date of Event
February 5, 2017
Report Date
February 6, 2017
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K042007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED INTERCONNECTION FLEX CABLE. THE FLEX CABLE WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A (B)(6) MALE PATIENT, THE DEVICE DISPLAYED A "CHECK PADS" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
141735 E SERIES DEFIBRILLATOR E SERIES MKJ ZOLL MEDICAL CORPORATION E SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 71 YR