DIRECT DRIVE DISPOSABLE CLIP APPLIER 10MM 3/BOX
Report
- Report Number
- 2027111-2017-00204
- Event Type
- Malfunction
- Date Received
- February 24, 2017
- Date of Event
- August 29, 2014
- Report Date
- February 24, 2017
- Manufacturer
- APPLIED MEDICAL RESOURCES
- Product Code
- FZP
- UDI-DI
- 00607915117306
- PMA / PMN Number
- 19
- Removal / Correction Number
- Z-1621-2016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
ON MARCH 18, 2016, APPLIED MEDICAL ISSUED A VOLUNTARY CLASS II RECALL OF THE CA090 DIRECT DRIVE CLIP APPLIER DUE TO INCREASED CUSTOMER FEEDBACK INDICATING INCONSISTENT CLIP APPLICATION, WHICH HAS THE POTENTIAL TO LEAD TO UNOCCLUDED VESSELS. AN INTERNAL CORRECTIVE AND PREVENTATIVE ACTION (CAPA) HAS BEEN INITIATED TO CONDUCT A THOROUGH INVESTIGATION AND IMPLEMENT APPROPRIATE CORRECTIVE ACTIONS. FDA HAS ISSUED RECALL NUMBER Z-1621-2016 FOR THIS RECALL. THIS REPORT IS BEING FILED AS A RESULT OF A RE-REVIEW OF APPLIED MEDICAL COMPLAINTS RECEIVED BETWEEN JUNE 1, 2014 AND MAY 31, 2016. THIS RETROSPECTIVE REVIEW WAS ASSOCIATED WITH A QUALITY MANAGEMENT SYSTEM (QMS) COMPLIANCE ACTION PLAN DEVELOPED AND PRESENTED TO FDA TO ADDRESS AN APRIL 10, 2015 WARNING LETTER. APPLIED MEDICAL HAS REVISED ITS MDR REPORTING CRITERIA TO BE MORE CONSERVATIVE AND HAS IMPROVED COMPLAINT HANDLING AND MDR REPORTING PROCESSES. THE REVIEWS ENSURED THAT RECENT REPORTABLE EVENTS WERE APPROPRIATELY IDENTIFIED AND REPORTED TO THE DESIGNATED REGULATORY AUTHORITY(IES). THIS REPORT, WHICH REPRESENTS THE INITIAL AND FINAL REPORTS COMBINED, IS BEING SUBMITTED BASED ON THE FINDINGS OF THAT RETROSPECTIVE REVIEW. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THIS REPORT WAS SUBMITTED, THEN A SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA.
LAP CHOLE - "DR. (B)(6) SAID THE CLIP APPLIER JAMMED AND WOULD ONLY DELIVER CLIP IF FORCED (1223891). THE SECOND CLIP APPLIER JAMMED AND THE CLIP WAS DELIVERED SIDEWAYS (1223891). THE THIRD CLIP APPLIER WAS NOT USED IN THE PATIENT AND JAMMED BEFORE IT WAS USED. (1224485)." PATIENT STATUS - UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 138820 | DIRECT DRIVE DISPOSABLE CLIP APPLIER 10MM 3/BOX | FZP | FZP | APPLIED MEDICAL RESOURCES | 19 | 1224485 | 00607915117306 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10MM CLIP APPLIER |