SUPRAMID BLACK 4/0 (1.5) 45CM HS21
Report
- Report Number
- 3003639970-2017-00071
- Event Type
- Malfunction
- Date Received
- February 24, 2017
- Report Date
- March 22, 2017
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- GAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR ANY OTHER TECHNOLOGICAL CHARACTERISTICS WITH A MEDICAL DEVICE THAT IS REGISTERED WITHIN THE U.S. AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN SURGICAL S.A. (MANUFACTURER). EXEMPTION NUMBER: E2014012. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.
COUNTRY OF COMPLAINT: (B)(6). IT WAS REPORTED THAT THE PACKAGE WAS LABELED (B)(4) BUT THE INSIDE PRODUCT WAS LABELED (B)(4). THE WRONG PRODUCT WAS INSIDE THE BOX.
SAMPLES RECEIVED: 1 UNOPENED BOX (36 POUCHES). ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS BATCH. MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS BATCH. THERE ARE NO UNITS IN STOCK. RECEIVED 1 CLOSED BOX. THE BOX LABEL AND PRODUCT IS SUPRAMID BLACK 4/0 (1.5) 45CM HS21 (CODE C0712256 AND BATCH 616466) BUT THE BOX THAT CONTAINS THE PRODUCT IS A DAFILON PRE-PRINTED BOX. THIS MISTAKE TOOK PLACE IN PRODUCTION AREA WHEN PACKING THE PRODUCT INTO THE BOXES, THE OPERATOR TOOK A DAFILON BOX INSTEAD OF A SUPRAMID'S. CORRECTIVE/PREVENTIVE ACTIONS: THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE. NEVERTHELESS, THE STORAGE LOCATIONS OF THE PRE-PRINTED BOXES WILL BE REARRANGING TO AVOID THIS DEFECT AND BIN IDENTIFICATION LABELING WILL BE CHANGED TO IMPROVE VISIBILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 138653 | SUPRAMID BLACK 4/0 (1.5) 45CM HS21 | SUTURES | GAR | B.BRAUN SURGICAL SA | C0712256 | 616466V004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |