FDA Adverse Event Injury Summary report: N

STRATA MR SHUNT VALVE

MDR report key: 6348661 · Received February 20, 2017

Report

Report Number
MW5067984
Event Type
Injury
Date Received
February 20, 2017
Date of Event
January 23, 2017
Report Date
February 20, 2017
Manufacturer
MEDTRONIC
Product Code
JXG
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

PATIENT HAD A STRATA MR VALVE INSERTED WITH A PRIMARY/HEW SHUNT INSERTION ON (B)(6) 2016. SHE REPRESENTED WITH SYMPTOMS OF SHUNT MALFUNCTION AND WAS REVISED ON (B)(6) 2017. WHEN PATIENT WAS EVALUATED INTRAOPERATIVELY, SHE WAS FOUND TO HAVE SLUGGISH/SLOW DRAINAGE WHEN ASSESSED WITH THE MANOMETER THROUGH THE VALVE ALTHOUGH SHE WAS SET AT A SHUNT VALVE SETTING OF 1.5. THE VALVE WAS NOTED TO ALSO HAVE LIMITED TO NO DRAINAGE THROUGH THE MANOMETER ONCE THE FLUID RAN DOWN TO A LEVEL OF ABOUT 25CM H20. THE VALVE WAS EXPLANTED AND REPLACED WITH A FIXED PRESSURE VALVE. DATES OF USE: (B)(6) 2016-(B)(6) 2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126518 STRATA MR SHUNT VALVE SHUNT JXG MEDTRONIC

Patients

Seq Age Sex Outcome Treatment
1 7 MO Hospitalization| R