FDA Adverse Event
Injury
Summary report: N
STRATA MR SHUNT VALVE
MDR report key: 6348661
·
Received February 20, 2017
Report
- Report Number
- MW5067984
- Event Type
- Injury
- Date Received
- February 20, 2017
- Date of Event
- January 23, 2017
- Report Date
- February 20, 2017
- Manufacturer
- MEDTRONIC
- Product Code
- JXG
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
PATIENT HAD A STRATA MR VALVE INSERTED WITH A PRIMARY/HEW SHUNT INSERTION ON (B)(6) 2016. SHE REPRESENTED WITH SYMPTOMS OF SHUNT MALFUNCTION AND WAS REVISED ON (B)(6) 2017. WHEN PATIENT WAS EVALUATED INTRAOPERATIVELY, SHE WAS FOUND TO HAVE SLUGGISH/SLOW DRAINAGE WHEN ASSESSED WITH THE MANOMETER THROUGH THE VALVE ALTHOUGH SHE WAS SET AT A SHUNT VALVE SETTING OF 1.5. THE VALVE WAS NOTED TO ALSO HAVE LIMITED TO NO DRAINAGE THROUGH THE MANOMETER ONCE THE FLUID RAN DOWN TO A LEVEL OF ABOUT 25CM H20. THE VALVE WAS EXPLANTED AND REPLACED WITH A FIXED PRESSURE VALVE. DATES OF USE: (B)(6) 2016-(B)(6) 2017.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126518 | STRATA MR SHUNT VALVE | SHUNT | JXG | MEDTRONIC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 MO | Hospitalization| R |