FDA Adverse Event Injury Summary report: N

SICAT CLASSICGUIDE

MDR report key: 6348615 · Received February 22, 2017

Report

Report Number
3006098230-2017-00003
Event Type
Injury
Date Received
February 22, 2017
Date of Event
January 23, 2017
Report Date
February 21, 2017
Manufacturer
SICAT GMBH & CO. KG
Product Code
NDP
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS DOES NOT GIVE ANY INDICATION FOR A MALFUNCTION OR FAULT OF THE SICAT CLASSICGUIDE AS MANUFACTURED BY SICAT. THE GUIDE IS REPRESENTING THE DENTISTS PLANNING AND WAS MANUFACTURED ACCORDING TO THE PLANNING AND PRESCRIPTION OF THE DENTIST. THE 3D X-RAY SCAN USED FOR IMPLANT PLANNING SEEMS TO SHOW SEVERE ARTEFACTS, COMPLICATING A PRECISE PLANNING OF THE IMPLANT POSITIONS BY THE DENTIST. IMPLANT #5 AND IMPLANT #10 HAVE NOT BEEN PLANNED AT OPTIMUM POSITIONS.

Description of Event or Problem · 1

THE REPORTING DENTIST HAS USED A SICAT SURGICAL GUIDE (SICAT CLASSICGUIDE) FOR PREPARING THE OSTEOTOMIES (DRILL HOLE FOR ACCOMMODATING A DENTAL IMPLANT) FOR FOUR DENTAL IMPLANTS OF TYPE "ZIMMER DENTAL TSVT MTX TSVTB8". THIS WAS AN EDENTULOUS PATIENT. THE OSTEOTOMY FOR IMPLANT ADA 5 ENDED UP IN THE INCORRECT POSITION. THE DENTIST HAD TO BONE GRAFT THIS AREA. THE IMPLANT AT ADA 10 WASN'T STABLE, SO THE DENTIST HAD TO BONE GRAFT THAT AREA AS WELL. THE DENTIST CONFIRMED THE FIT AND THE OTHER TWO IMPLANTS WERE JUST FINE. THE DENTIST WILL GO BACK TO THE PATIENTS IN A COUPLE MONTHS ONCE THE GRAFTS ARE HEALED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132294 SICAT CLASSICGUIDE SURGICAL GUIDE NDP SICAT GMBH & CO. KG

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention IMPLANT ZIMMER DENTAL TSVT MTX TSVTB8