FDA Adverse Event Injury Summary report: N

ANGIODYNAMICS

MDR report key: 6347649 · Received February 10, 2017

Report

Report Number
6347649
Event Type
Injury
Date Received
February 10, 2017
Date of Event
January 25, 2017
Report Date
February 7, 2017
Manufacturer
ANGIODYNAMICS
Product Code
LJT
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
IL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

ON (B)(6) 2017, PATIENT GRIMACED WHEN NURSE FLUSHED PORT-A-CATH. IMAGING REVEALED A FRACTURED CENTRAL VENOUS CHEST PORT CATHETER. CHEST PORT WAS PLACED VIA LEFT SUBCLAVIAN VEIN APPROACH IN 2013. PATIENT UNDERWENT AN ENDOVASCULAR RETRIEVAL OF INTRAVASCULAR FOREIGN BODY, CENTRAL VENOUS CHEST PORT CATHETER REMOVAL. A SCOUT RADIOGRAPH DEMONSTRATES FRACTURE OF AN EXISTING LEFT SUBCLAVIAN VEIN APPROACH CENTRAL VENOUS CHEST PORT CATHETER. THE PROXIMAL PORTION OF THE CATHETER REMAINS ATTACHED TO THE CHEST PORT, WITH FRACTURE VISUALIZED AT THE LEVEL OF THE INFERIOR CORTEX OF THE CLAVICLE. THE FREE, DISTAL FRAGMENT OF THE PORT CATHETER COURSES FROM THE LEFT BRACHIOCEPHALIC VEIN TO A RIGHT VENTRICLE. CONCLUSION: SUCCESSFUL ENDOVASCULAR RETRIEVAL OF FRACTURED CHEST PORT CATHETER FRAGMENT. SUCCESSFUL REMOVAL OF LEFT SUBCLAVIAN VEIN APPROACH CHEST PORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107234 ANGIODYNAMICS SMALL PORT LJT ANGIODYNAMICS 606289

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention