FDA Adverse Event Malfunction Summary report: N

VHM SERIES VARIABLE HEIGHT MOUNT (WS01)

MDR report key: 6345378 · Received February 21, 2017

Report

Report Number
6345378
Event Type
Malfunction
Date Received
February 21, 2017
Date of Event
November 25, 2016
Report Date
January 30, 2017
Manufacturer
GCX CORPORATION
Product Code
FOX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

MONITOR DROPPED FROM LOCKED POSITION AND HIT STAFF MEMBER'S HEAD. THE VHM ARTICULATING ARM CONNECTED TO THE PHILIPS MONITOR LOCK LEVER RELEASED DUE TO A INTERNAL GAS SPRING WHICH HAD LOST ITS GAS CHARGE. IN CASES WHERE THE ARM IS LEFT UNUSED AND UNCHECKED FOR MANY MONTHS, GRADUAL LOSS OF PRESSURE MAY NOT BE NOTICED UNTIL THE LOCKING LEVER IS RELEASED. MANUFACTURER RESPONSE FOR ARTICULATING MOUNT ARM, VHM ARTICULATING MOUNT ARM FOR PHILIPS MONITOR (PER SITE REPORTER). GCX BULLETIN HAD BEEN SENT OUT TO THE IDENTIFIED OWNERS. PRIMARY CHILDRENS' HOSPITAL CLINICAL ENGINEERING HAD NOT RECEIVED NOTICE. PHILIPS HAD INSTALLED THIS MOUNT WITH THE MONITORS THAT PRIMARY HAD PURCHASED FROM THEM, SO PRIMARY WAS NOT IDENTIFIED AS A OWNER BUT PHILIPS WHO HAD PURCHASED FOR INSTALLATION OF THEIR MONITORS IN THE UNITS. COMMUNICATION WAS NOT PASSED ON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129285 VHM SERIES VARIABLE HEIGHT MOUNT (WS01) STAND, INFUSION FOX GCX CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 Other