FDA Adverse Event Injury Summary report: N

PRODIGY AUTOCODE

MDR report key: 6342057 · Received February 18, 2017

Report

Report Number
3005862821-2017-00017
Event Type
Injury
Date Received
February 18, 2017
Date of Event
January 20, 2017
Report Date
January 20, 2017
Manufacturer
OK BIOTECH CO., LTD.
Product Code
NBW
PMA / PMN Number
K073118
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. WE TESTED THE STANDBY CURRENT OF RETURNED METER, THE RESULT WAS 13.1¿A. THE CRITERIA IS <55¿A. PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. BECAUSE PATIENT DID NOT RETURN HIS STRIPS, WE TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND IN HOUSE STRIPS (STRIP LOT NUMBER: D160526-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 58/63 MG/DL, FOR LEVEL HIGH WERE 249/240 MG/DL. THE REQUEST CONTROL SOLUTION RANGES ARE: LEVEL LOW 30~80 MG/DL; LEVEL HIGH 190~290 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.

Description of Event or Problem · 1

IT WAS REPORTED THAT MEDICAL ATTENTION WAS SOUGHT ON (B)(6) 2007 AT 3:00AM AFTER THE END USER WAS RECEIVING INCONSISTENT READINGS FROM THE PRODIGY DIABETES GLUCOSE METER. THE END USER'S BLOOD GLUCOSE DURING THE MEDICAL EVENT WAS 110 MG/DL. THE END USER WAS INCOHERENT AND UNRESPONSIVE AND THE PARAMEDICS WERE CALLED. THE PARAMEDICS PERFORMED A BLOOD GLUCOSE TEST WITH THEIR METER WITH A RESULT OF 39 MG/DL AND ADMINISTERED AN IV OF FLUID TO ASSIST WITH STABILIZING THEIR GLUCOSE LEVELS. NO FURTHER INFORMATION WAS PROVIDED IN RELATIONS TO THE MEDICAL EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124786 PRODIGY AUTOCODE BLOOD GLUCOSE MONITORING DEVICE NBW OK BIOTECH CO., LTD. 51900

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention ASPIRIN| BUMEX| COREG| DIGOXIN| IMBUR| LORATADINE| MAGNESIUM OXIDE| MYCOPHENOLATE| PRAZOSIN| PREDNISONE| PROBIOTIC| PROGRAF| SINGULAR| SYNTHROID| WARFARIN