FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
MDR report key: 6341098
·
Received February 17, 2017
Report
- Report Number
- 3015876-2017-00206
- Event Type
- Malfunction
- Date Received
- February 17, 2017
- Date of Event
- January 24, 2017
- Report Date
- February 17, 2017
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K142430
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED ISSUE. PHYSIO OBSERVED THAT THE KEP NUT WHICH CONNECTS THE NEGATIVE TERMINAL OF THE POWER PCB CABLE TO BATTERY WELL # 1 WAS LOOSE. PHYSIO THEN TIGHTENED THE LOOSE KEP NUT AND NO FURTHER POWER DISCREPANCIES WERE OBSERVED. AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF MULTIPLE TIMES. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 122499 | LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 15 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |