LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 6341098 · Received February 17, 2017

Report

Report Number
3015876-2017-00206
Event Type
Malfunction
Date Received
February 17, 2017
Date of Event
January 24, 2017
Report Date
February 17, 2017
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K142430
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED ISSUE. PHYSIO OBSERVED THAT THE KEP NUT WHICH CONNECTS THE NEGATIVE TERMINAL OF THE POWER PCB CABLE TO BATTERY WELL # 1 WAS LOOSE. PHYSIO THEN TIGHTENED THE LOOSE KEP NUT AND NO FURTHER POWER DISCREPANCIES WERE OBSERVED. AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF MULTIPLE TIMES. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.