COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING
Report
- Report Number
- 0001825034-2017-00721
- Event Type
- Injury
- Date Received
- February 17, 2017
- Date of Event
- January 25, 2017
- Report Date
- February 15, 2017
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- PAO
- PMA / PMN Number
- PK080642
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. IN THE PACKAGE INSERT UNDER POSSIBLE ADVERSE EFFECTS, ITEM 10 STATES: FATIGUE FRACTURE OF COMPONENT CAN OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE WEIGHT. MEDICAL PRODUCT - COMP PRIMARY STEM 16MM STD CATALOG #: 113656 LOT #: 110350, COMP LOCKING SCREW CATALOG #: 180501 LOT #: 710430, COMP NON-LCKNG SCREW CATALOG #: 180509 LOT #: 999860, COMP RVRS SHDR GLEN BSPLT CATALOG #: 115330 LOT #: 628500, COMP NON-LCKNG SCREW CATALOG #: 180508 LOT #: 979470, COMP RVS CNTRL SCR CATALOG #: 115381 LOT #: 345270, STD HMRL BRNG CATALOG #: XL-115366 LOT #: 340340, VERSA DIAL/COMPREHENSIVE STD T PR ADAPTOR CATALOG #: 118001 LOT #: 182190, COMPREHENSIVE REVERSE SHOULDER GLENOSPHERE CATALOG #: 115326 LOT #: 484540 THIS IS 1 OF 5 REPORTS BEING FILED FOR THE SAME PATIENT (REFERENCE 1825034-2017-00721 - 00725).
IT IS REPORTED THAT THE PATIENT WAS REVISED DUE TO IMPLANT FRACTURE APPROXIMATELY 4 YEARS POST-IMPLANTATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 122845 | COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING | PROSTHESIS, SHOULDER | PAO | BIOMET ORTHOPEDICS | N/A | 157340 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R | SEE H10 SUMMARY |