MODULAR EBS KNEE JOINT
Report
- Report Number
- 9616494-2017-00015
- Event Type
- Malfunction
- Date Received
- February 17, 2017
- Report Date
- February 17, 2017
- Manufacturer
- OTTO BOCK HEALTHCARE GMBH
- Product Code
- ISY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE COMPLAINT NOTIFICATION ASSOCIATED WITH THIS DEVICE DESCRIBED THAT SCREWS BECAME LOOSE AND THE UPPER SECTION IS DISLOCATED FROM THE LOWER SECTION WHICH HAS CAUSED THE USER TO FALL. BASED ON THE INFORMATION PROVIDED THERE WERE NO SERIOUS INJURIES SUSTAINED AS A RESULT OF THE FAILURE. THE EVALUATION OF THIS SPECIFIC DEVICE WAS CONDUCTED AT THE MANUFACTURER'S SERVICE CENTER ON FEBRUARY 3RD, 2017. WE CAN CONFIRM THAT THE BOLTS IN THE POSTERIOR UPPER PART ARE LOOSE. THE FRONT FRAME IS BENT DUE TO FURTHER USAGE OF THE DEVICE WITH LOOSENED BOLTS. WE ALSO FOUND THAT THE CONNECTION ROD OF THE HYDRAULIC IS BROKEN. AN EXACT REASON FOR THE LOOSE BOLTS AND THE BROKEN CONNECTION ROD COULD NOT BE DETERMINED. THEREFORE, ADDITIONAL DATA REGARDING PATIENT AND USAGE OF THE DEVICE WERE REQUESTED. THESE DATA DID NOT PROVIDE ANY NEWS WHICH WOULD ALLOW US TO COME TO A MEANINGFUL RESULT. THE LOOSE BOLTS COULD BE THE REASON FOR THE FALL. NO SERIOUS INJURY OCCURRED DUE TO THIS FAILURE, BUT IT COULD HAVE LED TO PATIENT INJURY IF THE MALFUNCTION WERE TO RECUR.
KNEE JOINT WAS SENT IN FOR REPAIR. CUSTOMER STATED THAT THE UPPER SCREWS HAVE LOOSENED AND THE UPPER SECTION IS DISLOCATED FROM THE LOWER SECTION WHICH HAS CAUSED THE USER TO FALL. BASED ON THE INFORMATION PROVIDED THERE WERE NO SERIOUS INJURIES SUSTAINED AS A RESULT OF THE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 122575 | MODULAR EBS KNEE JOINT | MODULAR EBS KNEE JOINT | ISY | OTTO BOCK HEALTHCARE GMBH | 3R60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |