FDA Adverse Event Malfunction Summary report: N

ZENITH TX2 TAA ENDOVASCULAR GRAFT PROXIMAL COMPONENT

MDR report key: 6338508 · Received February 17, 2017

Report

Report Number
3002808486-2017-00508
Event Type
Malfunction
Date Received
February 17, 2017
Date of Event
January 22, 2013
Report Date
January 22, 2013
Manufacturer
WILLIAM COOK EUROPE
Product Code
MIH
UDI-DI
10827002526973
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER (B)(4). SIMILAR TO DEVICE UNDER 510(K) P070016. SUMMARY OF INVESTIGATIONAL FINDINGS: AN INVESTIGATION WAS PERFORMED BASED ON THE AVAILABLE INFORMATION, BUT UNFORTUNATELY IT WAS NOT POSSIBLE TO CONCLUDE AN ROOT CAUSE FOR THIS EVENT. IT SEEMS HOWEVER LIKELY THAT THE OBSERVED ADVANCEMENT DIFFICULTIES ARE RELATED TO PATIENT TORTUOUS ANATOMY. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.

Description of Event or Problem · 1

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE ACCESS ROUTE (LEFT) WAS SLIGHTLY TORTUOUS AND THE AORTA WAS TORTUOUS. THE PHYSICIAN ATTEMPTED TO ADVANCE THE DELIVERY SYSTEM TO THE TARGET SITE BUT FAILED DUE TO TORTUOUS AORTA. HE REPLACED WITH ANOTHER ONE TO COMPLETE THE PROCEDURE. PATIENT OUTCOME: THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
120964 ZENITH TX2 TAA ENDOVASCULAR GRAFT PROXIMAL COMPONENT MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH WILLIAM COOK EUROPE 10827002526973

Patients

Seq Age Sex Outcome Treatment
1