FDA Adverse Event Injury Summary report: N

MELODY TRANSCATHER PULMONARY VALVE

MDR report key: 6335591 · Received February 16, 2017

Report

Report Number
2025587-2017-00257
Event Type
Injury
Date Received
February 16, 2017
Date of Event
August 18, 2016
Report Date
January 17, 2017
Manufacturer
MEDTRONIC HEART VALVES DIVISION
Product Code
NPV
PMA / PMN Number
P140017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CITATION: SOSNOWSKI C ET AL. HYBRID PULMONARY ARTERY PLICATION FOLLOWED BY TRANSCATHETER PULMONARY VALVE REPLACEMENT: COMPARISON WITH SURGICAL PVR. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS 2016; 88: 804¿810. EARLIEST DATE OF PUBLISH USED FOR EVENT DATE. NO UNIQUE DEVICE IDENTIFIER (SERIAL/LOT) NUMBERS WERE PROVIDED; WITHOUT THIS INFORMATION IT COULD NOT BE DETERMINED WHETHER THESE OBSERVATIONS HAVE BEEN PREVIOUSLY REPORTED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION VIA LITERATURE REGARDING HYBRID PULMONARY VALVE REPLACEMENT (PVR) COMBINING SURGICAL PLICATION OF THE MAIN PULMONARY ARTERY FOLLOWED BY TRANSCATHETER PVR. ALL DATA WERE COLLECTED FROM A SINGLE CENTER BETWEEN JUNE 2012 AND JUNE 2015. THE STUDY POPULATION INCLUDED 21 PATIENTS (PREDOMINANTLY MALE; MEAN AGE 31 YEARS), 8 OF WHICH WERE IMPLANTED WITH MEDTRONIC MELODY VALVE (SERIAL NUMBERS NOT PROVIDED). AMONG ALL PATIENTS ADVERSE EVENTS INCLUDED: ONE VALVE DISLODGEMENT AND THREE MILD PULMONARY REGURGITATION. THE VALVE DISLODGEMENT OCCURRED DURING REMOVAL OF DELIVERY CATHETER SYSTEM (DCS) FOLLOWING VALVE DEPLOYMENT WHERE THE ¿CARROT¿ OF THE DCS BECAME CAUGHT ON THE VALVE STENT. VIGOROUS REMOVAL EFFORTS RESULTED IN VALVE MOVEMENT AWAY FROM THE INITIAL IMPLANT SITE. THE VALVE COULD NOT BE OPTIMALLY REPOSITIONED, THEREFORE THE PROCEDURE WAS CONVERTED TO SURGICAL PVR. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
119383 MELODY TRANSCATHER PULMONARY VALVE PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED NPV MEDTRONIC HEART VALVES DIVISION PB 10

Patients

Seq Age Sex Outcome Treatment
1 31 YR Required Intervention