FDA Adverse Event Malfunction Summary report: N

AUTO SUTURE PREMIUM MULTIFIRE TA 30-V3

MDR report key: 63319 · Received January 3, 1997

Report

Report Number
1219930-1997-00025
Event Type
Malfunction
Date Received
January 3, 1997
Date of Event
November 26, 1996
Report Date
December 5, 1996
Manufacturer
UNITED STATES SURGICAL CORP.
Product Code
GAG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE SUBJECT DEVICE REVEALED THAT THE DISTAL END OF THE "#2 LATCHING ROD" LIFTED AND BECAME DISENGAGED FROM THE SUPPORT BLOCK. INTERMITTENTLY, DURING APPROXIMATION OF THE INSTRUMENT, THE LATCH CAM ADVANCED THE ROD SLIGHTLY FORWARD WHICH RESULTED IN BINDING OF THE ROD BETWEEN THE CAM BRIDGE AND THE SUPPORT BLOCK. THIS INTERMITTENT BINDING PREVENTED THE INSTRUMENT FROM OPENING UPON ACTIVATION OF THE APPROXIMATING BUTTON. AS CORRECTIVE ACTION, PROCESS MODIFICATIONS WERE IMPLEMENTED TO MECHANICALLY PREVENT THE ROD FROM LIFTING AND ELIMINATE FURTHER OCCURRENCE OF THIS CONDITION.

Description of Event or Problem · 1

DURING A PNEUMONECTOMY PROCEDURE, THE INSTRUMENT WAS DIFFICULT TO OPEN. THE SURGEON USED ANOTHER INSTRUMENT TO COMPLETE THE PROCEDURE. NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTO SUTURE PREMIUM MULTIFIRE TA 30-V3 DISPOSABLE STAPLER GAG UNITED STATES SURGICAL CORP. NA N6D98

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN
2