IOLMASTER 500
Report
- Report Number
- 9615030-2017-00002
- Event Type
- Injury
- Date Received
- February 10, 2017
- Date of Event
- January 5, 2017
- Report Date
- August 3, 2017
- Manufacturer
- CARL ZEISS MEDITEC AG (JENA)
- Product Code
- HJO
- PMA / PMN Number
- K122418
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
THE MANUFACTURER REVIEWED THE DATA LOGS AND THE MEASUREMENT AND IOL CALCULATION PRINTOUTS FROM THE IOLMASTER. NO ABNORMALITIES WERE FOUND. THE MANUFACTURER PERFORMED AN IN-HOUSE INVESTIGATION OF THE DEVICE. THE RIGHT AND LEFT TEST EYE CALIBRATION VALUES WERE FOUND TO BE OUT OF TOLERANCE. HOWEVER, THE SLIGHT DEVIATION OF 0.14MM IN ANTERIOR CHAMBER DEPTH MEASUREMENT CANNOT EXPLAIN THE SURGICAL OUTCOME, AS THE DEVIATION WOULD ONLY LEAD TO A DIFFERENCE OF LESS THAN 0.5D IN OUTCOME. A NUMBER OF FACTORS NOT RELATED TO THE IOLMASTER MAY HAVE INFLUENCED THE SURGICAL OUTCOME. THE RISK OF MEASURING ERRORS IS DESCRIBED IN DETAIL ON PAGE 7 OF THE USER MANUAL (000000-1692-983-DOKS-US-170413). ON PAGES 26-30 OF THE MANUAL, THE USER IS GIVEN DIRECTIONS FOR CHECKING THE MEASUREMENT FUNCTIONS OF THE DEVICE PRIOR TO EACH USE. (B)(4).
THE HEALTH CARE PROFESSIONAL REPORTED THE FOLLOWING: THE OS POST REFRACTIVE OUTCOME AFTER A CATARACT SURGERY WITH AN INTRAOCULAR LENS (IOL) IMPLANTATION DIFFERED +2.25 DIOPTERS FROM THE TARGET REFRACTION. THE ALCON ACRYSOF (B)(4) LENS WAS USED. THE HEALTHCARE PROFESSIONAL INFORMED THAT A RE-OP IS PLANNED WHERE AN ADDITIONAL PIGGYBACK LENS WILL BE IMPLANTED. THE IOLMASTER WAS USED FOR THE ORIGINAL BIOMETRY MEASUREMENTS AND IOL CALCULATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 107147 | IOLMASTER 500 | BIOMICROSCOPE, SLIT-LAMP, AC-POWERED | HJO | CARL ZEISS MEDITEC AG (JENA) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |