FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 6323536 · Received February 10, 2017

Report

Report Number
1000113657-2017-00256
Event Type
Malfunction
Date Received
February 10, 2017
Date of Event
January 13, 2017
Report Date
February 16, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292006068
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT # (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE. (BATHROOM) TEST STRIP UDI# (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER AND RETURNED TEST STRIPS EVALUATED WITH NO DEFECTS FOUND. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER'S TEST STRIP HAD POOR STORAGE .(BATHROOM). (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 51 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 144 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 170 MG/DL USING TRUEMETRIX METER AND 163 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE BATHROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 01/25/2018 AND OPEN VIAL DATE IS 12/13/2016. CUSTOMER STATED SHE JUST DISCONTINUED TAKING GLYBURIDE 5MG/ ONCE A DAY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): THE 83MG/DL (B)(6) 2016 05:32 AM FASTING:YES, THE 81MG/DL (B)(6) 2016 05:41 AM FASTING:YES, THE 63MG/DL (B)(6) 2016 06:53 AM FASTING:YES, THE 59MG/DL (B)(6) 2016 05:58 AM FASTING:YES, THE 51MG/DL (B)(6) 2016 05:57 AM FASTING:YES. MEMORY CONCERNS: 51MG/DL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 51 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 144 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 170 MG/DL USING TRUEMETRIX METER AND 163 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE BATHROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 01/25/2018 AND OPEN VIAL DATE IS 12/13/2016. CUSTOMER STATED SHE JUST DISCONTINUED TAKING GLYBURIDE 5MG/ ONCE A DAY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): (B)(6). MEMORY CONCERNS:51MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104662 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1895 00021292006068

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY