FDA Adverse Event Malfunction Summary report: N

HEMOTHERM 400 CE

MDR report key: 6316722 · Received February 9, 2017

Report

Report Number
1516825-2017-00001
Event Type
Malfunction
Date Received
February 9, 2017
Date of Event
December 29, 2016
Report Date
February 9, 2017
Manufacturer
CINCINNATI SUB-ZERO, LLC.
Product Code
DWC
PMA / PMN Number
K122813
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CINCINNATI SUB-ZERO TECHNICAL SUPPORT RECOMMENDED SENDING A TECHNICIAN TO THE ACCOUNT. THE TECHNICIAN FOUND THE PUMP MALFUNCTIONING DUE TO AN INOPERABLE POWER SUPPLY BOARD. PER THE CSZ SERVICE MANUAL, THE HEMOTHERM MUST BE SERVICED AND/OR PREVENTIVE MAINTENANCE MUST BE PERFORMED AT SPECIFIC INTERVALS AS OUTLINED IN THE MANUAL. IMPROPER REPAIR AND INADEQUATE MAINTENANCE CAN RESULT IN DAMAGE TO THE HEMOTHERM SYSTEM AND PATIENT INJURY. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR DEVICES. THE TECHNICIAN REPLACED THE POWER SUPPLY BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

CINCINNATI SUB-ZERO RECEIVED A REPORT FROM THE ACCOUNT STATING THE PUMP STOPPED WORKING DURING RE-WARMING. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96312 HEMOTHERM 400 CE CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER DWC CINCINNATI SUB-ZERO, LLC. 86022

Patients

Seq Age Sex Outcome Treatment
1