HEMOTHERM 400 CE
Report
- Report Number
- 1516825-2017-00001
- Event Type
- Malfunction
- Date Received
- February 9, 2017
- Date of Event
- December 29, 2016
- Report Date
- February 9, 2017
- Manufacturer
- CINCINNATI SUB-ZERO, LLC.
- Product Code
- DWC
- PMA / PMN Number
- K122813
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
CINCINNATI SUB-ZERO TECHNICAL SUPPORT RECOMMENDED SENDING A TECHNICIAN TO THE ACCOUNT. THE TECHNICIAN FOUND THE PUMP MALFUNCTIONING DUE TO AN INOPERABLE POWER SUPPLY BOARD. PER THE CSZ SERVICE MANUAL, THE HEMOTHERM MUST BE SERVICED AND/OR PREVENTIVE MAINTENANCE MUST BE PERFORMED AT SPECIFIC INTERVALS AS OUTLINED IN THE MANUAL. IMPROPER REPAIR AND INADEQUATE MAINTENANCE CAN RESULT IN DAMAGE TO THE HEMOTHERM SYSTEM AND PATIENT INJURY. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR DEVICES. THE TECHNICIAN REPLACED THE POWER SUPPLY BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
CINCINNATI SUB-ZERO RECEIVED A REPORT FROM THE ACCOUNT STATING THE PUMP STOPPED WORKING DURING RE-WARMING. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96312 | HEMOTHERM 400 CE | CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER | DWC | CINCINNATI SUB-ZERO, LLC. | 86022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |