ZENITH TX2 TAA ENDOVASCULAR GRAFT PROXIMAL COMPONENT
Report
- Report Number
- 3002808486-2017-00375
- Event Type
- Death
- Date Received
- February 9, 2017
- Date of Event
- February 24, 2014
- Report Date
- February 24, 2014
- Manufacturer
- WILLIAM COOK EUROPE
- Product Code
- MIH
- UDI-DI
- 10827002526942
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). SIMILAR TO DEVICE UNDER 510(K): P070016. SUMMARY OF INVESTIGATIONAL FINDINGS: THE INVESTIGATION OF THIS COMPLAINT WAS BASED ON THE REPORTED INFORMATION ALONG WITH ONE IMAGE OF A PORTABLE CHEST X-RAY. ACCORDING TO THE IMAGE, AN AORTIC RUPTURE WITH A LARGE HEMATOMA WAS EVIDENT AT THE TIME OF THIS X-RAY. IT IS HOWEVER NOT POSSIBLE TO DETERMINE THE CAUSE BASED ON ONE SINGLE LOW QUALITY IMAGE. NOTHING INDICATES THAT PRODUCT WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: AN (B)(6) YEAR OLD MALE PATIENT UNDERWENT TAA REPAIR. THE PATIENT'S ANATOMICAL FORM WAS SUITABLE FOR ENDOVASCULAR REPAIR. THE ACCESS VESSEL WAS THE RIGHT EXTERNAL ILIAC ARTERY AND ZTEG-2P-38-152-PF WAS PLACED FIRST AND ZTEG-2D-42-144-PF WAS PLACED NEXT AND THE PROCEDURE WAS COMPLETED. WHEN THE PATIENT WAS GOING BACK TO THE WARD AFTER COMING OUT FROM UNDER ANESTHESIA, HE COMPLAINED A CHEST PAIN. THEN THE BLOOD PRESSURE DECREASED TO 50 FROM 90, THEN HE WENT INTO CARDIAC ARREST. PCPS WAS USED WITH CARDIAC MASSAGE, HOWEVER THE PATIENT DECEASED. ADDITIONAL INFORMATION RECEIVED (B)(6) 2014: AN AUTOPSY WAS NOT CONDUCTED. A X-RAY IMAGE WAS PROVIDED. THE PHYSICIAN COMMENTED AS "THE CAUSE OF THE DISSECTION WAS REALLY UNKNOWN SINCE IT HAD NOT CONFIRMED DURING THE PROCEDURE AND THE CONDITION SUDDENLY CHANGED AFTER THE INCISION WAS CLOSED. ONLY POSSIBLE CAUSE OF THE COULD BE A TOUCH UP BY BALLOON, BUT IT OCCURRED TOO LATE FROM THE TOUCH UP." THE VESSEL CONDITION WAS NOT FRAGILE. ALTHOUGH IT WAS SLIGHTLY SHAGGY AND TORTUOUS, SO THE PHYSICIAN USED A WIRE GUIDE WITH CAUTION AND NO INJURY BY THE WIRE GUIDE WAS CONFIRMED. PATIENT OUTCOME: THE PATIENT DECEASED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 100165 | ZENITH TX2 TAA ENDOVASCULAR GRAFT PROXIMAL COMPONENT | MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | WILLIAM COOK EUROPE | 10827002526942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 | Death |