FDA Adverse Event Injury Summary report: N

MASTERLOC CEMENTLESS TI COATED STEM SIZE 11 LAT

MDR report key: 6310742 · Received February 8, 2017

Report

Report Number
3005180920-2017-00024
Event Type
Injury
Date Received
February 8, 2017
Date of Event
January 6, 2017
Report Date
February 8, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
PMA / PMN Number
K151531
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEWS PERFORMED ON 07 FEBRUARY 2017. LOT 162773: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08 JUNE 2016. EXPIRATION DATE: 2021-05-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MECTACER HEAD BIOLOX OPTION Ø 28, CODE 01.29.230H, LOT. 167065 (K131518) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23 NOVEMBER 2016. EXPIRATION DATE: 2021-11-14. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MECTACER SLEEVE SIZE M, CODE 01.29.241A, LOT. 153578 (K131518) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 28 DECEMBER 2015. EXPIRATION DATE: 2020-12-07. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. DOUBLE MOBILITY HC LINER Ø 58/28, CODE 01.26.2858MHC, LOT. 157126 (K092265) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18 FEBRUARY 2016. EXPIRATION DATE: 2021-02-06. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. DEVICE NOT AVAILABLE.

Description of Event or Problem · 1

THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE INFECTION IS CONFIRMED AS STAPH. THE SURGEON REVISED THE HEAD, SLEEVE, LINER AND STEM. THE SURGEON LEFT THE CUP IN PLACE. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95367 MASTERLOC CEMENTLESS TI COATED STEM SIZE 11 LAT CEMENTLESS FEMORAL STEM LZO MEDACTA INTERNATIONAL SA 162773

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention