FDA Adverse Event Malfunction Summary report: N

MODULAR EBS KNEE JOINT

MDR report key: 6310538 · Received February 8, 2017

Report

Report Number
9616494-2017-00012
Event Type
Malfunction
Date Received
February 8, 2017
Report Date
January 12, 2017
Manufacturer
OTTO BOCK HEALTHCARE GMBH
Product Code
ISY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT NOTIFICATION ASSOCIATED WITH THIS DEVICE DESCRIBED THAT THE PROXIMAL HINGE OF THE KNEE JOINT BROKE AND A SCREW WAS MISSING. THE EVALUATION OF THIS SPECIFIC DEVICE WAS CONDUCTED AT THE MANUFACTURER'S SERVICE CENTER ON FEBRUARY 6TH, 2017. WE CAN CONFIRM THAT ONE BOLT IN THE POSTERIOR UPPER PART IS LOST. FURTHER USAGE OF THE KNEE JOINT WITH LOST BOLT LED TO DAMAGES AT FRONT AND REAR FRAME. THE LINKAGE (CUSTOMER SAYS HINGE) TOUCHES THE LOWER PART OF THE JOINT, BUT IT IS NOT BROKEN. AN EXACT REASON FOR THE LOST BOLT COULD NOT BE DETERMINED. NO FALL OR SERIOUS INJURY OCCURRED DUE TO THIS FAILURE, BUT IT COULD HAVE LED TO PATIENT INJURY IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

KNEE JOINT WAS SENT IN FOR REPAIR. CUSTOMER STATED THAT THE PROXIMAL HINGE OF THE KNEE JOINT WAS BROKEN AND A SCREW IS MISSING. THE PATIENT DID NOT FALL, NO SERIOUS INJURY OCCURRED DUE TO THIS FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95238 MODULAR EBS KNEE JOINT MODULAR EBS KNEE JOINT ISY OTTO BOCK HEALTHCARE GMBH 3R60

Patients

Seq Age Sex Outcome Treatment
1