DIMENSION®
Report
- Report Number
- 2517506-2017-00056
- Event Type
- Malfunction
- Date Received
- February 6, 2017
- Date of Event
- November 22, 2016
- Report Date
- February 6, 2017
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- JIF
- PMA / PMN Number
- K123320
- Removal / Correction Number
- 2517506-12/27/2016-004-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER (CCC). THE CAUSE OF THE DISCORDANT AMM RESULTS IS UNKNOWN. THE CUSTOMER RECALIBRATED THE INSTRUMENT WITH A NEW AMM LOT EA7322 AND SET NEW RANGES FOR THE QC. THE CUSTOMER RERAN THE QC AND THE RESULTS OBTAINED WERE WITHIN THE EXPECTED RANGES. SIEMENS HEALTHCARE DIAGNOSTICS IS CONDUCTING A RECALL FOR THE DIMENSION® AMMONIA FLEX® REAGENT CARTRIDGE/AMM DF119 (SMN # (B)(4)) KIT LOTS (FB7152, EB7180, BA7194, EA7223, BA7250) AND DIMENSION VISTA® AMMONIA FLEX® REAGENT CARTRIDGE / AMM K3119 (SMN #(B)(4)) KIT LOTS (16187BE, 16225BB,16265AB). SIEMENS HEALTHCARE DIAGNOSTICS HAS DETERMINED THAT DIMENSION AMM (DF119) FLEX REAGENT CARTRIDGE LOTS FB7152, EB7180, BA7194, EA7223, BA7250 AND DIMENSION VISTA AMM (K3119) FLEX REAGENT CARTRIDGE FLEX LOTS 16187BE, 16225BB, 16265AB DO NOT MEET THE 60-DAY CALIBRATION INTERVAL CLAIM DUE TO REAGENT INSTABILITY AND RESULTS MAY SHOW AN ABNORMAL ASSAY. THESE LOTS MAY EXHIBIT ACCURACY SHIFTS FOR PATIENT AND/OR QUALITY CONTROL RESULTS; WHICH MAY CAUSE LABORATORIES TO RECALIBRATE MORE FREQUENTLY THAN THE 60-DAY CLAIM IN THE INSTRUCTIONS FOR USE (IFU). AN URGENT FIELD SAFETY NOTICES (UFSN'S) DC17-01.A.OUS.DM AND DC 17.01A.OUS.DMV WERE SENT TO OUS CUSTOMERS AND AN URGENT MEDICAL DEVICE RECALLS (UMDR'S) DC17-01.A.US.DM AND DC 17.01A.US.DMV WERE SENT TO US CUSTOMERS IN DECEMBER 2016. THE UFSN AND UMDR INFORMS THE CUSTOMERS TO DISCONTINUE USE OF AND DISCARD THE AFFECTED KIT LOT. THE CUSTOMER HAS STATED THAT THEY RECEIVED THE UMDR LETTER WHICH WAS MAILED AFTER THE INCIDENT OCCURRED AT THEIR LAB. SIEMENS RECOMMENDS USING AN ALTERNATE LOT OF DIMENSION OR DIMENSION VISTA AMM SIEMENS IS CURRENTLY INVESTIGATING THE ROOT CAUSE OF THIS ISSUE.
DISCORDANT LOW AMMONIA (AMM) RESULTS WERE OBTAINED ON QC AND A PATIENT SAMPLE ON THE DIMENSION EXL SYSTEM. THE PATIENT RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE QC WAS RERUN AFTER RECALIBRATION AND HIGHER RESULTS WITHIN LABORATORY RANGES WERE OBTAINED. THE PATIENT SAMPLE WAS NOT RETESTED. THERE IS NO INDICATION THAT PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF THE DISCORDANT LOW AMM RESULTS. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DISCORDANT LOW AMM RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 86678 | DIMENSION® | DIMENSION® AMM AMMONIA FLEX® REAGENT CARTRIDGE | JIF | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | FB7152 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |