AMISTEM C, CEMENTED STEM SIZE 0 STD
Report
- Report Number
- 3005180920-2017-00014
- Event Type
- Injury
- Date Received
- February 3, 2017
- Date of Event
- January 4, 2017
- Report Date
- April 27, 2017
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- PMA / PMN Number
- K103189
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 04 APRIL 2017, THE R&D PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED ITEM AND COMMENTED AS FOLLOWS: THE AMISTEM C WAS ANALYZED AND NO PARTICULAR AND EVIDENT SIGNS WERE NOTICED, EXCEPT FOR SOME EVIDENT AND SIGNIFICANT SCRATCHES AND SIGNS ON THE NECK SURFACE, MOST PROBABLY OCCURRED DURING THE REVISION OF THE STEM. THE DM LINER WAS IN GOOD CONDITIONS, WHILE THE CERAMIC BALL HEAD SHOWED SOME SIGNS ON THE RIM, DUE TO THE EXTRACTION HITS. FROM THE RECEIVED PIECES IT WAS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT.
BATCH REVIEW PERFORMED ON 31 JANUARY 2017. LOT 112648: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 AUGUST 2011. EXPIRATION DATE: 2016-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ON 17 JANUARY 2017 THE MEDICAL AFFAIRS DIRECTOR PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: FEMORAL STEM MOBILIZATION IN HYBRID THA 1,5 YEARS AFTER PRIMARY OPERATION. ONLY PRE-REVISION X-RAYS ARE AVAILABLE, AND THE STEM LOOKS SIGNIFICANTLY TILTED IN VARUS POSITION. IT IS CONCEIVABLE THAT FOR UNKNOWN REASONS THE STEM HAD A VARUS TILT SINCE THE ORIGINAL TREATMENT, WITH POSSIBLE INTERRUPTION OF CEMENT MANTLE IN DISTAL-LATERAL POSITION, WHICH MAY HAVE CONTRIBUTED TO SECONDARY LOOSENING. WITH THE INFORMATION AVAILABLE, THIS DOES NOT SEEM TO BE A PROBLEM CAUSED BY A FAULTY IMPLANT. DEVICE NOT YET RECEIVED.
THE PATIENT CAME IN COMPLAINING OF INSTABILITY. THE SURGEON CONFIRMED THE STEM WAS LOOSE. THE CAUSE OF THE LOOSENING IS UNKNOWN. THE PATIENT DID NOT EXPERIENCE ANY TRAUMA. THE SURGEON REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 84000 | AMISTEM C, CEMENTED STEM SIZE 0 STD | CEMENTED FEMORAL STEM | JDI | MEDACTA INTERNATIONAL SA | 112648 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |