21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER
Report
- Report Number
- 1024879-2017-00009
- Event Type
- Malfunction
- Date Received
- February 2, 2017
- Date of Event
- January 17, 2017
- Report Date
- March 13, 2017
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- FMI
- PMA / PMN Number
- K982541
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- OTHER
Narratives
RESULTS: A SAMPLE IS NOT AVAILABLE FOR EVALUATION, HOWEVER THE CUSTOMER PROVIDED THREE PHOTOS OF THE AFFECTED DEVICE. A PHOTO INSPECTION SHOWED AN ECLIPSE UNIT WITH THE COLLAR DETACHED FROM THE HUB. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6203554. CONCLUSION: ALTHOUGH THE PHOTOS CONFIRMED THE REPORTED DEFECT, WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE. REMEDIAL ACTION REQUIRED: CAPA (B)(4) HAS BEEN INITIATED TO DOCUMENT THE INVESTIGATION PATH, ROOT CAUSE ANALYSIS AND REMEDIATION PLAN, INCLUDE CORRECTIVE AND PREVENTIVE ACTIONS FOR THE REPORTED DEFECT, HUB COLLAR SEPARATION.
(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT THE SAFETY MECHANISM OF A 21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER FELL OFF OF THE NEEDLE WHEN IT WAS ATTACHED TO A HOLDER. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 78410 | 21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER | SAMPLE NEEDLE | FMI | BECTON, DICKINSON & CO., (BD) | 6203554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |