FDA Adverse Event Malfunction Summary report: N

21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER

MDR report key: 6298344 · Received February 2, 2017

Report

Report Number
1024879-2017-00009
Event Type
Malfunction
Date Received
February 2, 2017
Date of Event
January 17, 2017
Report Date
March 13, 2017
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
PMA / PMN Number
K982541
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: A SAMPLE IS NOT AVAILABLE FOR EVALUATION, HOWEVER THE CUSTOMER PROVIDED THREE PHOTOS OF THE AFFECTED DEVICE. A PHOTO INSPECTION SHOWED AN ECLIPSE UNIT WITH THE COLLAR DETACHED FROM THE HUB. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6203554. CONCLUSION: ALTHOUGH THE PHOTOS CONFIRMED THE REPORTED DEFECT, WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED AS BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE. REMEDIAL ACTION REQUIRED: CAPA (B)(4) HAS BEEN INITIATED TO DOCUMENT THE INVESTIGATION PATH, ROOT CAUSE ANALYSIS AND REMEDIATION PLAN, INCLUDE CORRECTIVE AND PREVENTIVE ACTIONS FOR THE REPORTED DEFECT, HUB COLLAR SEPARATION.

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SAFETY MECHANISM OF A 21 G X 1.25 IN BD ECLIPSE¿ BLOOD COLLECTION NEEDLE WITH LUER ADAPTER FELL OFF OF THE NEEDLE WHEN IT WAS ATTACHED TO A HOLDER. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.