FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 6298015 · Received February 2, 2017

Report

Report Number
1000113657-2017-00186
Event Type
Malfunction
Date Received
February 2, 2017
Date of Event
January 9, 2017
Report Date
February 2, 2017
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292006051
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE. TEST STRIP (B)(4).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 211, 169 AND 175 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 150 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2017, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED A TEST RESULTS OF 299 MG/DL AND 306 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 01/31/2018 AND OPEN VIAL DATE AT TIME OF CALL IS TWO WEEKS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
80951 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1909 00021292006051

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY