FDA Adverse Event Malfunction Summary report: N

RAD-5V

MDR report key: 6297090 · Received February 2, 2017

Report

Report Number
2031172-2017-00127
Event Type
Malfunction
Date Received
February 2, 2017
Date of Event
January 9, 2017
Manufacturer
MASIMO - 40 PARKER
Product Code
DQA
PMA / PMN Number
K033998
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED DEVICE WAS EVALUATED. DURING EVALUATION THE UNIT WAS ABLE TO POWER ON; HOWEVER, THE BOTTOM ROW OF LED DIGITS DOES NOT ILLUMINATE. COMPONENT C70 IS BROKEN OFF AND MISSING FROM THE INSTRUMENT BOARD. C70 IS REQUIRED FOR LED PWM DRIVING OF THE LOWER LED DIGITS. WITH C70 MISSING, THE BOTTOM ROW OF LEDS WILL NOT FUNCTION.

Additional Manufacturer Narrative · 1

ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 2ND ROW OF DISPLAY IS NOT WORKING. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
81569 RAD-5V OXIMETER DQA MASIMO - 40 PARKER 25199

Patients

Seq Age Sex Outcome Treatment
1