FDA Adverse Event Injury Summary report: N

CAREX DESIGNER OFFSET CANE - BLUE

MDR report key: 6297006 · Received February 2, 2017

Report

Report Number
3012316249-2017-00011
Event Type
Injury
Date Received
February 2, 2017
Date of Event
November 10, 2016
Report Date
October 6, 2020
Manufacturer
UNKNOWN
Product Code
KHY
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 0

THE DEVICE WAS RETURNED FOR INSPECTION. UPON INSPECTION, THE ITEM WAS FOUND TO HAVE BEEN OPENED AND USED. THE CUSTOMER STATED THAT THE STRAP BROKE ON THE CANE CAUSING HER TO FALL AND GET HURT. THE STRAP IS CONNECTED TO A PLASTIC SLEEVE THAT IS THEN SECURED INTO THE FRAME OF THE CANE. THE PLASTIC SLEEVE APPEARS TO BE LOCKED INTO PLACE WITH GLUE. THE GLUE APPEARS TO HAVE COME UNDONE, RESULTING IN THE PLASTIC SLEEVE AND STRAP DETACHING FROM THE UNIT.

Description of Event or Problem · 1

THIS CUSTOMER HAD EXPERIENCED A PROBLEM WITH THE BLUE OFFSET CANE. THIS CUSTOMER/END-USER CLAIMS THAT THE STRAP BROKE ON THE CANE, CAUSING HER TO FALL AND GET HURT. THE EVENT HAD OCCURRED IN THE END-USER'S HOME, ON LAMINATE FLOORING, COMING INTO HER HOUSE VIA ENTRYWAY. THE END-USER WAS ALREADY GOING TO THE ORTHOPEDIC DOCTOR FOR HER FOOT, AND WAS UNABLE TO USE CRUTCHES, SO SHE WAS INSTRUCTED TO USE A CANE. SHE PUT THE STRAP ON HER WRIST, AND REACHED OUT TO PUT HER HAND ON THE HANDLE AND THE STRAP BROKE, CAUSING HER TO FALL ON HER RIGHT SHOULDER. SHE DIDN'T REALIZE HER SHOULDER WAS HURT UNTIL A FEW DAYS LATER. SHE WENT BACK TO OUR CUSTOMER USING ONE OF THEIR ELECTRIC CHAIRS, AND SHOWED THE REP THAT THE STRAP BROKE AND SHE FELL. THE REP (FROM OUR CUSTOMER) PROVIDED HER WITH A NEW CANE AND TOOK THE BROKEN ONE. SHE WENT TO THE ORTHOPEDIC DOCTOR TO FOLLOW UP ON HER FOOT, AND SHE TOLD THE DOCTOR HER SHOULDER WAS HURTING. THE DOCTOR PRESCRIBED MEDICINE FIRST, AND THAT DIDN'T WORK, SO THEN THE DOCTOR GAVE HER A CORTISONE SHOT. SHE IS STILL NOT FEELING BETTER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
80089 CAREX DESIGNER OFFSET CANE - BLUE CANE KHY UNKNOWN FGA51500 BLUE UNAVAILABLE

Patients

Seq Age Sex Outcome Treatment
1 Other