FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA MP ICD

MDR report key: 6296500 · Received February 2, 2017

Report

Report Number
2938836-2017-13523
Event Type
Malfunction
Date Received
February 2, 2017
Date of Event
January 3, 2017
Report Date
January 3, 2017
Manufacturer
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN AN ASYMPTOMATIC PATIENT PRESENTED IN-CLINIC FOR A FOLLOW-UP, POST-PACED T-WAVE OVERSENSING ON THE VENTRICULAR CHANNEL WAS OBSERVED ON A STORED EGN. PROGRAMMING CHANGES WERE MADE. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
79529 QUADRA ASSURA MP ICD IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) CD3369-40Q A000025255

Patients

Seq Age Sex Outcome Treatment
1 68 YR