FDA Adverse Event
Other
Summary report: N
BARD/DAVOL SILICONE DRAIN
MDR report key: 62945
·
Received January 7, 1997
Report
- Report Number
- 1018233-1997-00001
- Event Type
- Other
- Date Received
- January 7, 1997
- Date of Event
- November 27, 1996
- Manufacturer
- C.R. BARD, INC.
- Product Code
- GCY
- Report Source
- Manufacturer report
Narratives
Additional Manufacturer Narrative · 1
BASED ON THE ADDITIONAL INFO REC'D ON 1/28/97, IT IS BELIEVED THAT AN INTERNAL SUTURE ERRONEOUSLY PIERCED THE DRAIN & WAS NOT VISUALIZED PRIOR TO AN ATTEMPTED REMOVAL OF THE DRAIN. RESISTANCE WAS NOTED DURING THE REMOVAL ATTEMPT & FORCE WAS APPLIED. THIS INFO WAS PROVIDED BY THE REPORTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BARD/DAVOL SILICONE DRAIN | WOULD DRAIN-HUBLESS | GCY | C.R. BARD, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | INFO RECEIVED THAT STATED IT WAS BELIEVED THAT AN| INTERNAL SUTURE PIERCED THE DRAIN. |