FDA Adverse Event Injury Summary report: N

ENDURANT II BIFURCATED STENT GRAFT

MDR report key: 6294187 · Received February 1, 2017

Report

Report Number
2953200-2017-00136
Event Type
Injury
Date Received
February 1, 2017
Date of Event
January 20, 2017
Report Date
January 20, 2017
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: SEX, DATE OF BIRTH. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

OTHER RELEVANT DEVICE(S): ETLW1616C93EJ, SN (B)(4).

Description of Event or Problem · 1

AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 4.4 CM IN DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT ONE WEEK POST ENDURANT IMPLANT PROCEDURE, CT SCAN IMAGES SHOWED THROMBOSIS WAS OCCLUDING THE ENDURANT STENT GRAFTS STARTING FROM RIGHT COMMON ILIAC ARTERY TO THE BIFURCATED PART OF THE MAIN BODY BIFURCATED STENT GRAFT SYSTEM. THE ENDURANT CONTRALATERAL LIMB STENT GRAFT WAS CRUSHED AT TERMINAL BY THE MAIN BODY IPSILATERAL LIMB, AND THROMBOEMBOLISM OCCURRED IN THE RIGHT LEG. DURING THE INITIAL PROCEDURE STENT GRAFTS WERE APPROXIMATELY 20 MM IN DIAMETER, THE FINAL ANGIOGRAM SHOWED THERE WAS NO ABNORMALITY IN THE BLOODSTREAM AND THEREFORE NO KISSING BALLOON TECHNIQUE (KBT) WAS PERFORMED. THE PHYSICIAN ELECTED TO PERFORM A THROMBECTOMY WITH A FOGARTY CATHETER AND TWO ENDURANT ILIAC LIMBS WERE IMPLANTED BOTH SIDES RESPECTIVELY STARTING FROM BELOW THE MAIN BODY BIFURCATION, AND KBT WAS THEN PERFORMED. THE EVENT WAS RESOLVED. PER THE PHYSICIAN THE CAUSE OF THE EVENT IS RELATED TO THE PATIENT ANATOMY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76300 ENDURANT II BIFURCATED STENT GRAFT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND ETBF2816C145EJ

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention