FDA Adverse Event Injury Summary report: N

HYBRID GLENOID BASE 4MM SMALL

MDR report key: 6292840 · Received February 1, 2017

Report

Report Number
0001825034-2017-00214
Event Type
Injury
Date Received
February 1, 2017
Date of Event
June 23, 2016
Report Date
June 21, 2017
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWS
PMA / PMN Number
PK060694
Removal / Correction Number
NI
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. CONCOMITANT PRODUCT: COMPR NANO HMRL PPS 36MM, US-115736 LOT #402830, VERSA-DIAL/COMP TI STD TAPER, CATALOG #118001 LOT #465300, VERSA-DIAL 42X21X43 HUM HEAD, CATALOG #113034 LOT #212290, PT HYBRID GLEN POST REGENEREX, CATALOG #PT-113950, LOT #989790. REPORTED EVENT WAS CONFIRMED BY REVIEW OF X-RAYS PROVIDED. X-RAY REVIEWER STATED "REPORTED EVENT "DISLOCATION/SUBLUXATION" IS CONFIRMED. ANTERIOR SUBLUXATION OF THE HUMERAL COMPONENT IS DEMONSTRATED, WHICH MAY BE ASSOCIATED WITH SUBSCAPULARIS RUPTURE OR ANTERIOR DELTOID DYSFUNCTION. DEVICE COMPONENTS APPEAR INTACT. RADIOLUCENT LINE AT THE GLENOID SCREW BONE INTERFACE MAY INDICATE GLENOID COMPONENT LOOSENING IF NEW OR PROGRESSIVE. A BONE SPUR IS PRESENT AT THE INFERIOR GLENOID RIM." DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION(S) IS/ARE REQUIRED. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT IS NUMBER 3 OF 4 MDRS FILED FOR THE SAME PATIENT (REFERENCE 0001825034-2017-00211/00212/00213/00214).

Description of Event or Problem · 1

PATIENT UNDERWENT AN INITIAL SHOULDER ARTHROPLASTY, AND SUBSEQUENTLY NOTED PAIN BEGINNING APPROXIMATELY THREE MONTHS POST-OPERATIVELY AND CONTINUING AT TWELVE MONTH POST-OPERATIVE FOLLOW-UP. THE PATIENT ADDITIONALLY REPORTED A LACK OF RANGE OF MOTION APPROXIMATELY THREE MONTHS POST-OPERATIVELY. SUBSEQUENTLY, PATIENT REPORTED SUBLUXATION AND INSTABILITY RELATED TO CONNECTIVE TISSUE LAXITY IN THE OPERATIVE SHOULDER APPROXIMATELY TWELVE MONTHS POST-OPERATIVELY. A NERVE DEFICIT WAS ALSO NOTED AT TWELVE MONTH FOLLOW-UP, HOWEVER REPORTEDLY RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75686 HYBRID GLENOID BASE 4MM SMALL PROTHESIS, SHOULDER KWS BIOMET ORTHOPEDICS NI 817120

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other