FDA Adverse Event Injury Summary report: N

M2A-MAGNUM MODULAR HEAD SZ 52MM

MDR report key: 6291019 · Received January 31, 2017

Report

Report Number
0001825034-2017-00304
Event Type
Injury
Date Received
January 31, 2017
Date of Event
January 8, 2013
Report Date
January 26, 2017
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. MEDICAL PRODUCT - BIOMET M2A-MAGNUM CUP ITEM #: US157858 LOT#: 600540; BIOMET TAPERLOC CATALOG#: 11-103207 LOT#: 061520.

Additional Manufacturer Narrative · 1

THE FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

Description of Event or Problem · 1

PATIENT REPORTED A RIGHT HIP REVISION APPROXIMATELY 4 YEARS POST-IMPLANTATION DUE TO UNSPECIFIED ALLEGATIONS. DURING THE PROCEDURE, THE FEMORAL HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED AND AN ACTIVE ARTICULATION BEARING WAS IMPLANTED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PATIENT'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

PATIENT REPORTED A RIGHT HIP REVISION APPROXIMATELY 4 YEARS POST-IMPLANTATION DUE TO UNSPECIFIED ALLEGATIONS. DURING THE PROCEDURE, THE FEMORAL HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED AND AN ACTIVE ARTICULATION BEARING WAS IMPLANTED. ADDITIONAL INFORMATION RECEIVED IN PATIENT'S MEDICAL RECORDS INDICATES THE PATIENT WAS REVISED DUE TO EVIDENCE OF A PSEUDOTUMOR AND AN ASEPTIC HIP. REVISION OPERATIVE REPORT NOTED EXCISION OF THE PSEUDOTUMOR, DEBRIDEMENT OF SCAR TISSUE AND DEBRIS, AND CORROSION AT THE HEAD AND TAPER JUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71816 M2A-MAGNUM MODULAR HEAD SZ 52MM PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 713170

Patients

Seq Age Sex Outcome Treatment
1 37 YR Hospitalization| R