I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2017-00006
- Event Type
- Malfunction
- Date Received
- January 27, 2017
- Date of Event
- January 18, 2017
- Report Date
- February 17, 2017
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT #(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT # 711811. THE INVESTIGATION WAS COMPLETED ON 02/09/2017. RETAIN PRODUCT WAS TESTED AND THE CUSTOMER COMPLAINT WAS NOT REPRODUCED. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.
ON (B)(6) 2017, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT SODIUM, POTTASSIUM & HEMATOCRIT ON A (B)(6) MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. CUSTOMER STATES THAT PRODUCT IS AVAILABLE FOR INVESTIGATION: METHOD DATE COLLECTED TESTED TCO2 NA K CI HCT: I-STAT 01/18/2017 1802 1810 31 111 3.3 >140 39; VISTA 01/18/2017 1802 1837 25 134 3.9 98 43; I-STAT 01/18/2017 1835 1835 24 134 3.8 96 47. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67163 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H16253A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |