FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 6284059 · Received January 27, 2017

Report

Report Number
2245578-2017-00006
Event Type
Malfunction
Date Received
January 27, 2017
Date of Event
January 18, 2017
Report Date
February 17, 2017
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

APOC INCIDENT #(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 1

APOC INCIDENT # 711811. THE INVESTIGATION WAS COMPLETED ON 02/09/2017. RETAIN PRODUCT WAS TESTED AND THE CUSTOMER COMPLAINT WAS NOT REPRODUCED. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.

Description of Event or Problem · 1

ON (B)(6) 2017, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT SODIUM, POTTASSIUM & HEMATOCRIT ON A (B)(6) MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. CUSTOMER STATES THAT PRODUCT IS AVAILABLE FOR INVESTIGATION: METHOD DATE COLLECTED TESTED TCO2 NA K CI HCT: I-STAT 01/18/2017 1802 1810 31 111 3.3 >140 39; VISTA 01/18/2017 1802 1837 25 134 3.9 98 43; I-STAT 01/18/2017 1835 1835 24 134 3.8 96 47. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67163 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H16253A

Patients

Seq Age Sex Outcome Treatment
1 46 YR