FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 6283503 · Received January 27, 2017

Report

Report Number
2954323-2017-00843
Event Type
Malfunction
Date Received
January 27, 2017
Date of Event
January 17, 2017
Report Date
January 29, 2018
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K092602
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

AN EXTENDED INVESTIGATION HAS BEEN CONDUCTED, WHICH INVOLVED A MANUFACTURING REVIEW (INCLUDING 5 YEARS OF DEVICE HISTORY RECORDS (DHRS)), PREVIOUS COMPLAINT AND CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATIONS CONDUCTED FOR BLANK SCREEN ISSUES, PROCESS FAILURE MODE EFFECTS ANALYSES (PFMEAS), DESIGN CONTROLS, AND DESIGN FAILURE MODE EFFECTS ANALYSES (DFMEAS), RISK MANAGEMENT REPORTS, RISK EVALUATIONS, AND LABEL COPY. THE INVESTIGATION DID NOT IDENTIFY ANY INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. IF THE PRODUCT IS RETURNED, AN INVESTIGATION WILL BE PERFORMED.

Additional Manufacturer Narrative · 1

UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

THIS SERVES AS A CORRECTION REPORT. BRAND NAME WAS INCORRECTLY IDENTIFIED AS FREESTYLE FREEDOM. THE CORRECT BRAND NAME FOR THE REPORTED METER SERIAL NUMBER IS (B)(4). BRAND NAME HAS BEEN UPDATED TO REFLECT THE CORRECTION. IN ADDITION PMA/510K# HAS BEEN UPDATED TO K092602 AND CATALOG NO HAS BEEN UPDATED TO 70914-70.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65742 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC 70914-70

Patients

Seq Age Sex Outcome Treatment
1 63 YR