FDA Adverse Event Malfunction Summary report: N

ROHS MICROSENSOR BASIC KIT

MDR report key: 6276019 · Received January 25, 2017

Report

Report Number
1226348-2017-10063
Event Type
Malfunction
Date Received
January 25, 2017
Date of Event
November 3, 2016
Manufacturer
CODMAN & SHURTLEFF, INC.
Product Code
GWM
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). 510 (K) # OF SIMILAR PRODUCT CODE OF 826631: K914479. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED FOR EVLAUTION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION BY THE SUPPLIER. A REVIEW OF QUALITY RECORDS FOUND THAT THE DEVICE MET ALL MANUFACTURING AND QUALITY TESTING/INSPECTION SPECIFICATIONS PRIOR TO DISTRIBUTION. EVALUATION OF THE RETURNED DEVICE FOUND THAT THERE WAS AN UNKNOWN FILM ON THE SENSOR AREA THAT IS NOT PART OF THE MANUFACTURING PROCESS. THE DEVICE PASSED ELECTRONIC, NOISE, LINEARITY/HYSTERESIS, AND SIGNAL DRIFT TESTS. BASED ON THEIR EVALUATION, THE SUPPLIER WAS NOT ABLE TO CONFIRM THE REPORTED ISSUE. THE SENSOR FUNCTIONED AS INTENDED. TRENDS WILL BE MONITORED FOR THIS OR SIMILAR COMPLAINTS. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.

Description of Event or Problem · 1

MICROSENSOR ONLY ALLOWS NEGATIVE ICP READINGS. ICU DEPT REMOVED MICROSENSOR AND PUT IN ALTERNATIVE UNIT OF 626631 AND READINGS WERE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59786 ROHS MICROSENSOR BASIC KIT DEVICE, INTRACRANIAL PRESSURE MONITORING GWM CODMAN & SHURTLEFF, INC. 880366

Patients

Seq Age Sex Outcome Treatment
1