FDA Adverse Event
Malfunction
Summary report: N
MICROSCAN RAPID POST INOCULUM BROTH
MDR report key: 627524
·
Received August 13, 2005
Report
- Report Number
- 2919016-2005-00012
- Event Type
- Malfunction
- Date Received
- August 13, 2005
- Date of Event
- December 22, 2004
- Report Date
- December 22, 2004
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- JSO
- Removal / Correction Number
- Z-0668-05
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
QC S. AUREUS ATCC 29213 AND CLINICAL ISOLATES VERSUS OXACILLIN WERE OUT HIGH (RESISTANT) WHEN USED WITH RAPID POSITIVE MIC LOT #2005-03-14. THIS ISSUE HAS BEEN ASSOCIATED WITH A DADE BEHRING CONDUCTED RECALL FOR MICROSCAN RAPID POS BROTH INOCOLUM BROTH, LOTS 050905A-1, 051605A-1, 051605A-1, 051705A-1, 052305A-1 AND 052805B-1 THAT TOOK PLACE IN 01/05.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSCAN RAPID POST INOCULUM BROTH | INOCULUM BROTH | JSO | DADE BEHRING, INC. | NA | MULTIPLE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |