SMARTSET MV 40G - EO
Report
- Report Number
- 1818910-2017-11296
- Event Type
- Malfunction
- Date Received
- January 24, 2017
- Date of Event
- January 17, 2017
- Report Date
- January 17, 2017
- Manufacturer
- DEPUY CMW 9610921
- Product Code
- LOD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. (B)(4).
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. IT HAS BEEN NOTED, THAT APPLYING TOO MUCH PRESSURE DURING THE SNAPPING OF THE AMPOULE SEAL COULD POTENTIALLY LEAD TO THE BREAKING OF GLASS INTO THE AMPOULE OR FLAKING GLASS PARTICLES. THE IFU (IFU-0163-146 REV 3) CONTAINS ADVICE TO BREAK THE AMPOULE AWAY FROM THE MIXING DEVICE TO PREVENT POSSIBLE CONTAMINATION FROM BROKEN AMPOULE GLASS. DVA-107020-FDE REV 5, LINES 101 AND 119 (BAR) WHICH STATES THE POTENTIAL HAZARD: ¿DAMAGED DEVICE. DAMAGED GLASS AMPOULE DUE TO INADEQUATE PROTECTION FROM SECONDARY PACKAGING.¿ A STUDY WAS CONDUCTED AND HAS BEEN DOCUMENTED AS FOLLOWS: ¿TRANSIT STUDY VERIFICATION HAS BEEN COMPLETED TO ENSURE THE PACKAGING SYSTEM IS SUITABLE FOR USE AND REMAINS INTEGRAL. REFERENCE: RR-ENG-101349-TR ISTA 3A TRANSIT STUDY FOR AMPOULE BLISTER AND CARTON.¿ LINE103(BAR) STATES ¿MMA IS A STRONG SOLVENT AND AT THE TIME OF DEVELOPMENT GLASS AMPOULES WERE THE MOST SUITABLE PACKAGING VESSEL FOR LONG TERM STORAGE¿ (REPORT REFERENCE R374). A LIKELY ROOT CAUSE FOR THIS FAILURE MODE IS THAT A HAIRLINE FRACTURE WAS PRESENT IN THE AMPOULE, WHICH COULD HAVE OCCURRED FROM THE ¿BUMPING¿ OF AMPOULES ALONG THE PRODUCTION LINE DURING THE LIQUID FILLING PROCESS. THIS POSSIBLE FRACTURE COULD HAVE BEEN AMPLIFIED DURING TRANSIT DUE TO TRANSIT STRESSES, WHICH COULD HAVE RESULTED IN THE AMPOULE BREAKING IN A WAY NOT SEEN BY THE SURGICAL TEAM PREVIOUSLY. PREVIOUS EFFECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO IMPROVE THE PROTECTION OF AMPOULES DURING PACKING AND TRANSIT TO THE CUSTOMER. COMPLAINTS RECEIVED FOR THIS ISSUE WILL CONTINUE TO BE MONITORED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
GLASS FRAGMENTS BROKE OFF OF MV MONOMER (LIQUID) GLASS VIAL. FRAGMENTS FELL INTO THE POWDER IN THE MIXING BOWL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57700 | SMARTSET MV 40G - EO | CEMENT / CEMENT ACCESSORY | LOD | DEPUY CMW 9610921 | 8374100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |