ROD, FIXATION, INTRAMEDUALLARY
Report
- Report Number
- 2520274-2017-10244
- Event Type
- Injury
- Date Received
- January 23, 2017
- Report Date
- January 16, 2017
- Manufacturer
- SYNTHES USA
- Product Code
- HSB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE USED FOR TREATMENT, NOT FOR DIAGNOSIS. DATE OF EVENT: FLORES, S; WOOLRIDGE, A; CAROOM, C; AND JENKINS, M. (2016) THE UTILITY OF THE TIP¿APEX DISTANCE IN PREDICTING AXIAL MIGRATION AND CUTOUT WITH THE TROCHANTERIC FIXATION NAIL SYSTEM HELICAL BLADE. J ORTHO TRAUM, VOL 30:6, E207-E211. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO DEVICE WAS RETURNED AND NO LOT NUMBER OR PART NUMBER WAS PROVIDED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THIS REPORT IS BEING FILED AFTER SUBSEQUENT REVIEW OF THE FOLLOWING LITERATURE ARTICLE: FLORES, S; WOOLRIDGE, A; CAROOM, C; AND JENKINS, M. (2016) THE UTILITY OF THE TIP¿APEX DISTANCE IN PREDICTING AXIAL MIGRATION AND CUTOUT WITH THE TROCHANTERIC FIXATION NAIL SYSTEM HELICAL BLADE. J ORTHO TRAUM, VOL 30:6, E207-E211. THE PURPOSE OF THE RETROSPECTIVE STUDY WAS TO EVALUATE THE MECHANICAL COMPLICATIONS ASSOCIATED WITH THE HELICAL BLADE DESIGNED IMPLANT OF THE TROCHANTERIC FIXATION NAIL (TFN) SYSTEM ((B)(4)), PARTICULARLY WITH REGARD TO AXIAL MIGRATION AND CUTOUT, AND FURTHERMORE TO DETERMINE THE UTILITY OF THE TIP-APEX DISTANCE (TAD) IN PREDICTING THIS COMPLICATION. FOUR HUNDRED FORTY-THREE PATIENTS SURGICALLY TREATED FOR PROXIMAL FEMUR FRACTURES WERE IDENTIFIED THROUGH A QUERY OF CURRENT PROCEDURAL TERMINOLOGY (CPT) CODE 27245 FROM JANUARY 2009 TO FEBRUARY 2013. A TOTAL OF 258 PATIENTS TREATED WITH THE TFN HELICAL BLADE SYSTEM WERE INCLUDED IN THE STUDY. ONE HUNDRED FIFTY-EIGHT PATIENTS WERE FEMALE AND A HUNDRED WERE MALE. THE MEAN AGE WAS 77 YEARS. AVERAGE FOLLOW-UP WAS 6.5 MONTHS USING THE ARBEITSGEMEINSCHAFT FUR OSTEOSYNTHESEFRAGEN/ ORTHOPEDIC TRAUMA ASSOCIATION (AO/OTA) CLASSIFICATION SYSTEM, 137 FRACTURES WERE STABLE AND 121 WERE UNSTABLE. THE AVERAGE TAD WAS 20.4 MM (RANGE 3.8¿41.56 MM). THE OVERALL RATE OF MECHANICAL COMPLICATIONS WAS 10.1%. AXIAL MIGRATION WITH PERFORATION OCCURRED IN 8 PATIENTS. AXIAL MIGRATION ONLY UP TO SUBCHONDRAL BONE WITHOUT PERFORATION OCCURRED IN 3 PATIENTS. OF THE 8 PATIENTS WITH FEMORAL HEAD AXIAL PERFORATIONS, 7 REQUIRED REVISION SURGERY, WITH 3 PATIENTS REQUIRING MULTIPLE REVISION PROCEDURE INCLUDING ONE PATIENT REQUIRING EVENTUAL TOTAL HIP ARTHROPLASTY AND ANOTHER PATIENT REQUIRING A GIRDLESTONE PROCEDURE. ONE PATIENT AND THEIR FAMILY ELECTED TO SIMPLY LIVE WITH LIMITED WEIGHT BEARING AND NOT PROCEED WITH ANY ADDITIONAL REVISION PROCEDURES. SYMPTOMATIC LATERAL MIGRATION OCCURRED IN 11 PATIENTS, ALTHOUGH ONLY 1 REQUIRED A REVISION PROCEDURE. THREE FRACTURES OCCURRED AT THE DISTAL TIP OF THE NAIL (1.1%). ONE FRACTURE FAILED IN VARUS COLLAPSE AND CUTOUT. A TAD <20 MM WAS ASSOCIATED WITH AN INCREASED FREQUENCY OF AXIAL MIGRATION AND CUTOUT WITH THE TFN HELICAL BLADE. THIS REPORT IS 14 OF 14 FOR (B)(4). THIS REPORT IS FOR AN UNKNOWN TFN SYSTEM AND REFERS TO THE SERIOUS INJURY OF 3 UNKNOWN PATIENTS WHO EXPERIENCED FRACTURES AT DISTAL TIP OF NAIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 53623 | ROD, FIXATION, INTRAMEDUALLARY | HSB | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |