FDA Adverse Event Malfunction Summary report: N

THERMASOLUTIONS, INC.

MDR report key: 6269558 · Received January 20, 2017

Report

Report Number
MW5067378
Event Type
Malfunction
Date Received
January 20, 2017
Date of Event
January 17, 2017
Report Date
January 19, 2017
Manufacturer
THERMASOLUTIONS, INC.
Product Code
MLW
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

TEMPERATURE SENSOR ERROR DURING SET-UP. VENDOR DISCOVERED CORROSION ON INTERNAL THERMOSTAT LOCATED IN WATER RESERVOIR. SURGERY COULD NOT BE COMPLETED DUE TO MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
49451 THERMASOLUTIONS, INC. THERMOCHEM HT-1000 MLW THERMASOLUTIONS, INC. MT-1000

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other