FDA Adverse Event Malfunction Summary report: N

MAXI TWIN

MDR report key: 6269382 · Received January 23, 2017

Report

Report Number
3007420694-2017-00016
Event Type
Malfunction
Date Received
January 23, 2017
Report Date
February 20, 2017
Manufacturer
ARJOHUNTLEIGH POLSKA SP. Z O.O.
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED UNDER EXEMPTION E2012070 BY THE MANUFACTURER ARJOHUNTLEIGH POLSKA SP. Z O.O. (REGISTRATION#3007420694) ON BEHALF OF THE IMPORTER ARJOHUNTLEIGH, INC. (AHUS) (REGISTRATION#1419652). ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSIONS FROM INVESTIGATION.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED UNDER EXEMPTION E2012070 BY THE MANUFACTURER ARJOHUNTLEIGH POLSKA SP. Z O.O. (REGISTRATION#3007420694) ON BEHALF OF THE IMPORTER ARJOHUNTLEIGH, INC. (AHUS) (REGISTRATION#1419652). AN INVESTIGATION WAS CARRIED OUT INTO THIS COMPLAINT. ARJOHUNTLEIGH RECEIVED A COMPLAINT WHERE IT WAS INDICATED THAT STITCHING INSIDE OF THE YELLOW LOOP OF THE SLING FAILED BASED ON THE INFORMATION RECEIVED. NO CONSEQUENCES TO RESIDENT WERE REPORTED. WHEN REVIEWING SIMILAR REPORTABLE EVENTS, WE HAVE FOUND A NUMBER OF CASES WITH SIMILAR FAULT DESCRIPTION (LOOP STITCHING INSIDE LOOP FAILED). THE TREND OBSERVED FOR REPORTABLE COMPLAINTS WITH THIS FAILURE MODE IS CURRENTLY CONSIDERED TO BE LOW AND STABLE. IT HAS BEEN ESTABLISHED THAT THE SLING WAS BEING USED FOR PATIENT HANDLING AT THE TIME OF THE EVENT. DURING OUR INVESTIGATION THE SLING WAS FOUND YELLOW LOOP STITCHING INSIDE LOOP FAILED AND WAS FOUND TO NOT HAVE BEEN TO SPECIFICATION. IT WAS REPORTED THAT MAXI TWIN LIFT DEVICE WAS USED WITH THIS SLING. NO MALFUCTIONS REGARDING LIFT WERE REPORTED. TESTS CARRIED OUT DURING THE DEVELOPMENT OF THE SLING, AND IN CURRENT PRODUCTION ON EVERY SLING MANUFACTURED, WOULD INDICATE THE STITCHING OF THE LOOPS MEETS THE OEM SPECIFICATION. A PROPER INSPECTION OF THE SLING SHOULD HAVE DETECTED THE FAILURE OF THIS SLING, ESPECIALLY SINCE ONE OF ATTACHMENT POINTS OF THE LOOP WAS BROKEN. THEREFORE, THE SLING SHOWING SIGNS OF UNSTITCHING, SHOULD BE WITHDRAWN AND REPLACED. AFTER REVIEWING THE COMPLAINT IT COMES FORWARD THAT THE LOOP BREAKAGE APPEARS MOST LIKELY TO OCCUR AS FOLLOWS: AS THE PRESSURE ON THE SLING LOOP, IN THE OPPOSITE DIRECTION OF THAT EXPERIENCED IN NORMAL USE WHICH CAN BE CAUSED BY THE CAREGIVER PULLING THE LOOPS THAT WAY BY HAND, OR, A MUCH HIGHER STRAIN, WHERE THE LOOP/SLING BECOMING CAUGHT IN AN OBSTRUCTION. THIS APPEARS TO BE AN INDICATION OF THE LOOP BEING BROKEN DUE TO THE APPLICATION OF OUTSIDE FORCE THAT CAUSES THE BREAK. SINCE THE LOOP BREAK IS INDICATED TO HAVE OCCURRED OUTSIDE OF USE, IT MAY HAVE BEEN CAUSED EVEN BY BECOMING STUCK DURING THE MECHANICAL WASHING OR DRYING PROCESS. AFTER THIS REVIEW, WE CAN STATE THE EVENT OUTCOME OF THE BREAKING OFF THE LOOP, IS NOT LIKELY TO HAPPEN WHEN FOLLOWING THE DEVICE LABELING OR THE INSTRUCTIONS FOR USE (IFU). THE SLING IS NOT LIKELY TO FAIL DURING THE INTENDED, CORRECT USE AS DESCRIBED IN THE IFU, BUT THAT A FAILURE CAN OCCUR DURING A USE ERROR. WE FIND IT LIKELY THAT THE LOOP BROKE DUE TO THE LOOP SUFFERING STRESS THAT IS NOT LIKELY TO BE ENCOUNTERED DURING ON LABEL USE. WE FIND THIS COMPLAINT TO BE REPORTABLE TO THE COMPETENT AUTHORITIES IN THE ABUNDANCE OF CAUTION.

Description of Event or Problem · 0

ON (B)(6) 2016 ARJOHUNTLEIGH RECEIVED A COMPLAINT WHERE IT WAS INDICATED THAT STITCHING INSIDE OF THE YELLOW LOOP OF THE SLING FAILED BASED ON THE PHOTOGRAPHIC EVIDENCES RECEIVED. NO INJURIES AND NO PATIENT INVOLVEMENT WAS REPORTED.

Description of Event or Problem · 0

ON 26 DEC 2016 ARJOHUNTLEIGH RECEIVED A COMPLAINT WHERE IT WAS INDICATED THAT STITCHING INSIDE OF THE YELLOW LOOP OF THE SLING FAILED BASED ON THE PHOTOGRAPHIC EVIDENCES RECEIVED. NO CONSEQUENCES TO RESIDENT WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
52800 MAXI TWIN LIFT, PATIENT, NON-AC-POWERED FSA ARJOHUNTLEIGH POLSKA SP. Z O.O. KTBB4BSA2WW

Patients

Seq Age Sex Outcome Treatment
1 Other