FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 6266591 · Received January 20, 2017

Report

Report Number
2245578-2017-00004
Event Type
Malfunction
Date Received
January 20, 2017
Date of Event
January 8, 2017
Report Date
February 10, 2017
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 01/31/2017. RETAIN PRODUCT AND CUSTOMER RETURNS WERE TESTED AND THE CUSTOMER COMPLAINT WAS NOT REPRODUCED.

Description of Event or Problem · 1

ON (B)(6) 2017, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT SODIUM, CHLORIDE & HEMATOCRIT ON A (B)(6) YEAR OLD FEMALE PATIENT PRESENTED IN THE ER WITH SHORTNESS OF BREATH. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. CUSTOMER STATES THAT PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. DATE: SAMPLE: TIME: NA: CI: HCT: (B)(6) 2017, A, 15:23, 110, >140, 31%. (B)(6) 2017, A, 15:29, 143, >105, 40%. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50525 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H16253A

Patients

Seq Age Sex Outcome Treatment
1 52 YR